SwePub
Sök i LIBRIS databas

  Utökad sökning

WFRF:(Abrahamsson Pernilla 1972 )
 

Sökning: WFRF:(Abrahamsson Pernilla 1972 ) > Optimised sample ha...

Optimised sample handling in association with use of the CMA 600 analyser

Abrahamsson, Pernilla, 1972- (författare)
Umeå universitet,Anestesiologi och intensivvård,Enheten för Anestesi och Intensivvård
Johansson, Göran (författare)
Umeå universitet,Anestesiologi och intensivvård
Åberg, Anna-Maja (författare)
Umeå universitet,Anestesiologi och intensivvård
visa fler...
Haney, Michael (författare)
Umeå universitet,Anestesiologi och intensivvård
Winsö, Ola (författare)
Umeå universitet,Anestesiologi och intensivvård
visa färre...
 (creator_code:org_t)
Elsevier, 2008
2008
Engelska.
Ingår i: Journal of Pharmaceutical and Biomedical Analysis. - : Elsevier. - 0731-7085 .- 1873-264X. ; 48:5, s. 940-945
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • A large degree of variability for batched analysis of serially collected microdialysis samples measured with the CMA 600 analyser has been described. This study was designed to identify sources of variability related to sample handling. Standard concentrations of four solutes were placed in microdialysis vials and then stored and analysed at intervals. Results were analysed for variability related to vial and cap type, duration and temperature of storage, centrifugation and re-analysis. The main results were that centrifugation of samples reduced variability. When a batch of 24 samples was analysed, the use of crimp caps reduced evaporation. Samples in glass vials with crimp caps could be stored in a refrigerator for up to 14 days without large variability in concentration compared to plastic vials which demonstrated variability already when stored for more than 1 day. We conclude that variability in microdialysis results can occur in relation to storage and analysis routines if routines are not optimised concerning evaporation. Centrifugation before analyses, glass vials with crimp caps even during frozen storage, and attention to minimal times for samples to be uncapped during analysis all contribute to minimise variability in the handling and analysis of microdialysis samples.

Nyckelord

Microdialysis; CMA 600; Microdialysis validation; Microdialysis sample; Microdialysis sample storage

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy