  • Nilsson, Kjell GUmeå universitet,Ortopedi (författare)

Assessment of horizontal laminar air flow instrument table for additional ultraclean space during surgery

  • Artikel/kapitelEngelska2010

Förlag, utgivningsår, omfång ...

  • Elsevier,2010
  • printrdacarrier

Nummerbeteckningar

  • LIBRIS-ID:oai:DiVA.org:umu-37546
  • https://urn.kb.se/resolve?urn=urn:nbn:se:umu:diva-37546URI
  • https://doi.org/10.1016/j.jhin.2010.05.016DOI

Kompletterande språkuppgifter

  • Språk:engelska
  • Sammanfattning på:engelska

Ingår i deldatabas

Klassifikation

  • Ämneskategori:ref swepub-contenttype
  • Ämneskategori:art swepub-publicationtype

Anmärkningar

  • The area in a vertical ultraclean laminar air flow (LAF) theatre is usually too small to accommodate all the equipment needed for major surgery. We investigated the addition of an instrument table supplied with fixed ultraclean LAF and placed alongside the existing main LAF unit, to determine its physical and bacteriological effect on the main unit. In phase 1, with two investigators but without a patient, smoke tests showed no intrusion of air from the table into the main unit and particle counts did not show any adverse effect on the main LAF unit. In phase 2, during patients undergoing two total knee replacements, the LAF table and a table without LAF were placed alongside the main LAF unit. The tables were subjected to the activity of an extra operating room (OR) nurse working from inside the main LAF vigorously simulating handling of instruments. During this activity, the >5 μm particle counts were 275/m3 at the instrument table with LAF and 8550/m3 at the table without LAF (P < 0.0001). Also, without the OR nurse activity, the particle counts, just inside the main unit and adjacent to the LAF table, were significantly reduced (P < 0.03–0.003). Sedimentation plates on the LAF table and in the main unit registered 22 and 25 cfu/m2/h respectively compared with 45 cfu/m2/h at the instrument table without LAF. In conclusion, the results from the smoke tests, particle counts and bacteriological evaluation showed that the additional instrument table supplied with LAF is efficient and can be safely used as an extension additional to a main OR LAF unit.

Ämnesord och genrebeteckningar

  • Air contamination
  • Colony count
  • Laminar air flow
  • Operating room
  • Particle count

Biuppslag (personer, institutioner, konferenser, titlar ...)

  • Lundholm, RolfUmeå universitet,Institutionen för klinisk mikrobiologi(Swepub:umu)rolu0012 (författare)
  • Friberg, SvenUmeå universitet,Ortopedi (författare)
  • Umeå universitetOrtopedi (creator_code:org_t)

Sammanhörande titlar

  • Ingår i:Journal of Hospital Infection: Elsevier76:3, s. 243-2460195-67011532-2939

Internetlänk

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Nilsson, Kjell G
Lundholm, Rolf
Friberg, Sven
Artiklar i publikationen
Journal of Hospi ...
Av lärosätet
Umeå universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy