In vitro testing of prevailing materials and initial clinical findings

Stegmayr, B. G. (författare)
Umeå universitet,Medicin
Forsberg, U. (författare)
Umeå universitet,Medicin
Stegmayr, C. J. (författare)
Umeå universitet,Medicin
visa fler...
Jonsson, P. (författare)
Umeå universitet,Medicin
visa färre...
 (creator_code:org_t)
2011
2011
Engelska.
Ingår i: International Journal of Artificial Organs. - 0391-3988 .- 1724-6040. ; 34:8, s. 636-636
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Objectives: During HD previous studies have shown that especially micro bubbles of air may pass the air detector. These studies focused to analyse in vitro if the air trap of various producers may contribute to the presence of micro embolic counts in the fluid that has passed the air trap detector. In addition another in vitro study analysed if the dialyzer by itself may contribute to these contaminations. In parallel a clinical study was performed to evaluate if findings during in vitro tests could be found to some extent during chronic hemodialysis. If such contamination was present how frequent this would be and would it be a greater risk at the start or the end of dialysis. Methods: A ultrasound probe was placed on the venous dialysis tube after the air detector and venous chamber (Hatteland, Norway). Results:The studies verified previous in vitro studies with micro embolic counts that pass the air trap without inducing an alarm. Fewer embolic signals were detected in the in vitro studies when using a high level of the fluid in the air trap and when using a wet dialyzers. In the clinical studies high counts were present both at the first period as well as the last period of hemodialysis. Conclusions: These studies verify the finding of embolic counts in various extents in various in vitro settings, but also verify the presence of such counts in the dialysis tubes after the air trap in clinical routine hemodialysis. Further studies are warranted to clarify how to prevent patients from such problems and to clarify if the counts are clinically relevant.

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Stegmayr, B. G.
Forsberg, U.
Stegmayr, C. J.
Jonsson, P.
Artiklar i publikationen
International Jo ...
Av lärosätet
Umeå universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy