SwePub
Sök i LIBRIS databas

  Utökad sökning

L773:1528 1132
 

Sökning: L773:1528 1132 > Wear of a 5 Megarad...

Wear of a 5 Megarad Cross-linked Polyethylene Liner : A 6-year RSA Study

Callary, Stuart A. (författare)
Campbell, David G. (författare)
Mercer, Graham (författare)
visa fler...
Nilsson, Kjell G. (författare)
Umeå universitet,Ortopedi
Field, John R. (författare)
visa färre...
 (creator_code:org_t)
Ovid Technologies (Wolters Kluwer Health), 2013
2013
Engelska.
Ingår i: Clinical Orthopaedics and Related Research. - : Ovid Technologies (Wolters Kluwer Health). - 0009-921X .- 1528-1132. ; 471:7, s. 2238-2244
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • One cross-linked polyethylene (XLPE) liner is manufactured using a lower dose of radiation, 5 Mrad, which may result in less cross-linking. The reported in vivo wear rate of this XLPE liner in patients undergoing THA has varied, and has included some patients in each reported cohort who had greater than 0.1 mm/year of wear, which is an historical threshold for osteolysis. Previous studies have measured wear on plain radiographs, an approach that has limited sensitivity. We therefore measured the amount and direction of wear at 6 years using Radiostereometric analysis (RSA) in patients who had THAs that included a cross-linked polyethylene liner manufactured using 5 Mrad radiation. We prospectively reviewed wear in 30 patients who underwent primary THAs with the same design of cross-linked acetabular liner and a 28-mm articulation. Tantalum markers were inserted during surgery and all patients had RSA radiographic examinations at 1 week, 6 months, 1, 2, and 6 years postoperatively. The mean proximal, two-dimensional (2-D) and three-dimensional (3-D) wear rates calculated between 1 year and 6 years were 0.014, 0.014, and 0.018 mm/per year, respectively. The direction of the head penetration recorded between 1 week and 6 years was in a proximal direction for all patients, proximolateral for 16 of 24 patients, and proximomedial for eight of 24 patients. The proximal, 2-D and 3-D wear of a XLPE liner produced using 5 Mrad of radiation was low but measurable by RSA after 6 years. No patients had proximal 2-D or 3-D wear rates exceeding 0.1 mm/year. Further followup is needed to evaluate the effect of XLPE wear particles on the development of long-term osteolysis.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Ortopedi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Orthopaedics (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Callary, Stuart ...
Campbell, David ...
Mercer, Graham
Nilsson, Kjell G ...
Field, John R.
Om ämnet
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Klinisk medicin
och Ortopedi
Artiklar i publikationen
Clinical Orthopa ...
Av lärosätet
Umeå universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy