A comparative study of extra- and intraluminal sigmoid colonic tonometry to detect colonic hypoperfusion after operation for abdominal aortic aneurysm

Djavani-Gidlund, Khatereh, 1967- (författare)
Uppsala universitet,Kärlkirurgi,Centrum för klinisk forskning, Gävleborg
Wanhainen, Anders (författare)
Uppsala universitet,Kärlkirurgi
Björck, Martin (författare)
Uppsala universitet,Kärlkirurgi
 (creator_code:org_t)
Elsevier BV, 2011
2011
Engelska.
Ingår i: European Journal of Vascular and Endovascular Surgery. - : Elsevier BV. - 1078-5884 .- 1532-2165. ; 42:3, s. 302-308
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Objectives: There is no ideal method to monitor colonic perfusion after abdominal aortic aneurysm (AAA) repair. The aim was to evaluate extraluminal sigmoid colon tonometry, comparing with the established intraluminal method. Methods: Eighteen patients were monitored with both methods, 10 after elective and eight after ruptured AAA repair. One tonometric catheter was placed inside the sigmoid colon (intraluminal) and another extraluminally in close contact with the serosa of the sigmoid colon (extraluminal). Intra- and extraluminal partial pressure of carbon dioxide (pCO2) were measured every 10 min during 48 h postoperatively, 1536 simultaneous measurements. Intraluminal pH (pHi) and extraluminal pH (pHe) were calculated, and intra-abdominal pressure (IAP) was measured, every 4 h. Colonic ischaemia was defined as pHi ≤ 7.1. Results: Mean pHi was 7.18 ± 0.11 and mean pHe was 7.28 ± 0.09. With a pHe cut-off value of ≤7.2, the sensitivity and specificity to detect colonic ischaemia were 95% and 95%, respectively. Accuracy was 95% and the positive and negative predictive values 0.80 and 0.99, respectively. The positive likelihood ratio was 19 and the negative likelihood ratio 0.05. Conclusion: Extraluminal tonometry may serve as a screening test: A pHe-value <7.2 indicates suspected colonic ischaemia, meriting further investigation. It was not able to evaluate the severity of ischaemia.

Nyckelord

Tonometry
CO2 partial pressure
Colonic ischaemia
Abdominal aortic aneurysm
Rupture
Intra-abdominal pressure
Intra-abdominal hypertension

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy