Comparison of seven infant continuous positive airway pressure systems using simulated neonatal breathing

Drevhammar, Thomas (författare)
Karolinska Institutet
Nilsson, Kjell (författare)
Zetterström, Henrik (författare)
Uppsala universitet,Anestesiologi och intensivvård
visa fler...
Jonsson, Baldvin (författare)
Karolinska Institutet
visa färre...
 (creator_code:org_t)
2012
2012
Engelska.
Ingår i: Pediatric Critical Care Medicine. - 1529-7535 .- 1947-3893. ; 13:2, s. E113-E119
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Objective: Continuous positive airway pressure is an established treatment for respiratory distress in neonates. Continuous positive airway pressure has been applied to infants using an array of devices. The aim of this experimental study was to investigate the characteristics of seven continuous positive airway pressure systems using simulated breath profiles from newborns. Design: Experimental in vitro study. Setting: Research laboratory in Sweden. Intervention: None. Measurements and Main Results: In vitro simulation of spontaneous neonatal breathing was achieved with a mechanical lung model. Simulation included two breath profiles, three levels of continuous positive airway pressure with and without short binasal prongs and different levels of constant leak. Pressure stability and imposed work of breathing were determined. Seven continuous positive airway pressure systems were tested. There were large differences in pressure stability and imposed work of breathing between tested continuous positive airway pressure systems. Neopuff and Medijet had the highest pressure instability and imposed work of breathing. Benveniste, Hamilton Universal (Arabella), and Bubble continuous positive airway pressure showed intermediate results. AirLife and Infant Flow had the lowest pressure instability and imposed work of breathing. AirLife and Infant Flow showed the least decrease in delivered pressure when challenged with constant leak. Conclusion: The seven tested continuous positive airway pressure systems showed large variations in pressure stability and imposed work of breathing. They also showed large differences in how well they maintain continuous positive airway pressure when exposed to leak. For most systems, imposed work of breathing increased with increasing continuous positive airway pressure level. The clinical importance of the difference in pressure stability is uncertain. Our results may facilitate the design of clinical studies examining the effect of pressure stability on outcome. 

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Pediatrik (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Pediatrics (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Anestesi och intensivvård (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Anesthesiology and Intensive Care (hsv//eng)

Nyckelord

continuous positive airway pressure
equipment design
infant
newborn
positive-pressure respiration/instrumentation
respiratory mechanics
work of breathing

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy