Equal clinical performance of a novel point-of-care cardiac troponin I (cTnI) assay with a commonly used high-sensitivity cTnI assay

Venge, Per (författare)
Uppsala universitet,Institutionen för medicinska vetenskaper
van Lippen, Lian (författare)
Philips Handheld Diagnost, Eindhoven, Netherlands.
Blaschke, Sabine (författare)
Univ Med Ctr Gottingen, Interdisciplinary Emergency Care Unit, Gottingen, Germany.
visa fler...
Christ, Michael (författare)
Luzerner Kantonsspital, Emergency Dept, Luzern, Switzerland.;Paracelsus Med Univ, Nuernberg Gen Hosp, Dept Emergency & Crit Care Med, Nurnberg, Germany.
Geier, Felicitas (författare)
Paracelsus Med Univ, Nuernberg Gen Hosp, Dept Emergency & Crit Care Med, Nurnberg, Germany.
Giannitsis, Evangelos (författare)
Heidelberg Univ, Med Klin 3, Heidelberg, Germany.
Hagström, Emil (författare)
Uppsala universitet,Uppsala kliniska forskningscentrum (UCR)
Hausfater, Pierre (författare)
Hop La Pitie Salpetriere, AP HP, Emergency Dept, Paris, France.;UPMC Univ Paris 06, Sorbonne Univ, GRC BIOSFAST 14, Paris, France.
Khellaf, Mehdi (författare)
Hop Henri Mondor, Emergency Dept, Creteil, France.
Mair, Johannes (författare)
Innsbruck Med Univ, Dept Internal Med Cardiol & Angiol 3, Innsbruck, Austria.
Pariente, David (författare)
Hop La Pitie Salpetriere, AP HP, Emergency Dept, Paris, France.
Scharnhorst, Volkher (författare)
Catharina Ziekenhuis Eindhoven, Clin Lab, Eindhoven, Netherlands.;Tech Univ Eindhoven, Eindhoven, Netherlands.
Semjonow, Veronique (författare)
Philips Handheld Diagnost, Eindhoven, Netherlands.
visa färre...
 (creator_code:org_t)
ELSEVIER SCIENCE BV, 2017
2017
Engelska.
Ingår i: Clinica Chimica Acta. - : ELSEVIER SCIENCE BV. - 0009-8981 .- 1873-3492. ; 469, s. 119-125
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background: Efficient rule-out of acute myocardial infarction (MI) facilitates early disposition of chest pain patients in emergency departments (ED). Point-of-care (POC) cardiac troponin (cTn) may improve patient throughput. We compared the diagnostic accuracy of a novel cTnI test (Minicare cTnI, Philips), with current POC cTnI (I-Stat, Abbott) and high-sensitivity central laboratory cTnI (hs-cTnI; Architect, Abbott) assays.Methods: The clinical performance of the assays were compared in samples from 450 patients from a previous clinical evaluation of Minicare cTnI.Results: Minicare cTnI correlated with Architect hs-cTnI (r(2) = 0.85, p < 0.0001) and I-Stat cTnI (r(2) = 0.93, p < 0.0001). Areas under the receiver operating characteristics curves were 0.87-0.91 at admission (p = ns) and 0.96-0.97 3 h after admission (p = ns). The negative predictive values (NPV) at admission were 95% ((92-97%, 95% CI) for Minicare cTnI and increased to 99% (97-100%) at 2-4 h, and similar to Architect hs-cTnI (98%, 96-100%), but higher than I-Stat cTnI (95%, 92-97%; p < 0.01). Negative likelihood ratios (LR) after 2-4 h were 0.06 (0.02-0.17, 95% CI) for Minicare cTnI, 0.11 (0.05-0.24) for Architect hs-cTnI (p = 0.02) and 0.28 (0.18-0.43) for I-Stat cTnI (p < 0.0001). The clinical concordances between Minicare cTnI and Architect hs-cTnI were 92% (admission) and 95% (2-4 h), with lower concordances between Minicare cTnI and I-Stat cTnI (83% and 78%, respectively; p = 0.007).Conclusions: The Minicare cTnI POC assay may become useful for prompt and safe ruling-out of AMI in ED patients with suspected AMI using a guideline supported 0/3 h sampling protocol.

Nyckelord

Cardiac troponin I
Acute myocardial infarction
Point-of-care
Emergency medicine

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy