SwePub
Sök i LIBRIS databas

  Utökad sökning

(WFRF:(Johansson Håkan 1979))
 

Sökning: (WFRF:(Johansson Håkan 1979)) > (2020-2023) > Validation of autom...

Validation of automated whole-body analysis of metabolic and morphological parameters from an integrated FDG-PET/MRI acquisition

Guglielmo, Priscilla (författare)
Uppsala universitet,Radiologi,University of Milan Bicocca, Milan, Italy
Ekström, Simon, 1991- (författare)
Uppsala universitet,Radiologi
Strand, Robin, 1978- (författare)
Uppsala universitet,Radiologi
visa fler...
Visvanathar, Robin (författare)
Uppsala universitet,Radiologi
Malmberg, Filip, 1980- (författare)
Uppsala universitet,Radiologi
Johansson, Emil (författare)
Uppsala universitet,Radiologi,GE Healthcare, Chicago, USA
Pereira, Maria J., 1981- (författare)
Uppsala universitet,Klinisk diabetologi och metabolism
Skrtic, S. (författare)
Carlsson, B. C. L. (författare)
Eriksson, Jan W. (författare)
Uppsala universitet,Klinisk diabetologi och metabolism
Ahlström, Håkan, 1953- (författare)
Uppsala universitet,Radiologi,Antaros Medical, Mölndal, Sweden
Kullberg, Joel, 1979- (författare)
Uppsala universitet,Radiologi,Antaros Medical, Mölndal, Sweden
visa färre...
 (creator_code:org_t)
2020-03-24
2020
Engelska.
Ingår i: Scientific Reports. - : Springer Science and Business Media LLC. - 2045-2322. ; 10:1
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Automated quantification of tissue morphology and tracer uptake in PET/MR images could streamline the analysis compared to traditional manual methods. To validate a single atlas image segmentation approach for automated assessment of tissue volume, fat content (FF) and glucose uptake (GU) from whole-body [18F]FDG-PET/MR images. Twelve subjects underwent whole-body [18F]FDG-PET/MRI during hyperinsulinemic-euglycemic clamp. Automated analysis of tissue volumes, FF and GU were achieved using image registration to a single atlas image with reference segmentations of 18 volume of interests (VOIs). Manual segmentations by an experienced radiologist were used as reference. Quantification accuracy was assessed with Dice scores, group comparisons and correlations. VOI Dice scores ranged from 0.93 to 0.32. Muscles, brain, VAT and liver showed the highest scores. Pancreas, large and small intestines demonstrated lower segmentation accuracy and poor correlations. Estimated tissue volumes differed significantly in 8 cases. Tissue FFs were often slightly but significantly overestimated. Satisfactory agreements were observed in most tissue GUs. Automated tissue identification and characterization using a single atlas segmentation performs well compared to manual segmentation in most tissues and will be valuable in future studies. In certain tissues, alternative quantification methods or improvements to the current approach is needed.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy