SwePub
Sök i LIBRIS databas

  Utökad sökning

WFRF:(Oldham M. A.)
 

Sökning: WFRF:(Oldham M. A.) > Report of AAPM Task...

Report of AAPM Task Group 155 : Megavoltage photon beam dosimetry in small fields and non-equilibrium conditions

Das, Indra J. (författare)
Northwestern Univ, Northwestern Mem Hosp, Feinberg Sch Med, Dept Radiat Oncol, Chicago, IL 60611 USA.
Francescon, Paolo (författare)
Osped Vicenza, Dept Radiat Oncol, Vicenza, Italy.
Moran, Jean M. (författare)
Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10021 USA.
visa fler...
Ahnesjö, Anders, 1953- (författare)
Uppsala universitet,Medicinsk strålningsvetenskap
Aspradakis, Maria M. (författare)
Cantonal Hosp Graubunden, Inst Radiat Oncol, Chur, Switzerland.
Cheng, Chee-Wai (författare)
Univ Hosp Cleveland, Dept Radiat Oncol, Med Ctr, Cleveland, OH 44106 USA.
Ding, George X. (författare)
Vanderbilt Univ, Sch Med, Dept Radiat Oncol, Nashville, TN 37212 USA.
Fenwick, John D. (författare)
Univ Liverpool, Inst Syst Mol & Integrat Biol, Mol & Clin Canc Med, Liverpool, Merseyside, England.
Huq, M. Saiful (författare)
Univ Pittsburgh, Sch Med, Dept Radiat Oncol, Pittsburgh, PA USA.;UPMC Hillman Canc Ctr, Pittsburgh, PA USA.
Oldham, Mark (författare)
Duke Univ, Dept Radiat Oncol, Med Ctr, Durham, NC USA.
Reft, Chester S. (författare)
Univ Chicago, Dept Radiat Oncol, Chicago, IL 60637 USA.
Sauer, Otto A. (författare)
Univ Wurzburg, Dept Radiat Oncol, Klin Strahlentherapie, Wurzburg, Germany.
visa färre...
Northwestern Univ, Northwestern Mem Hosp, Feinberg Sch Med, Dept Radiat Oncol, Chicago, IL 60611 USA Osped Vicenza, Dept Radiat Oncol, Vicenza, Italy. (creator_code:org_t)
2021-07-21
2021
Engelska.
Ingår i: Medical physics (Lancaster). - : John Wiley & Sons. - 0094-2405 .- 2473-4209. ; 48:10, s. E886-E921
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Small-field dosimetry used in advance treatment technologies poses challenges due to loss of lateral charged particle equilibrium (LCPE), occlusion of the primary photon source, and the limited choice of suitable radiation detectors. These challenges greatly influence dosimetric accuracy. Many high-profile radiation incidents have demonstrated a poor understanding of appropriate methodology for small-field dosimetry. These incidents are a cause for concern because the use of small fields in various specialized radiation treatment techniques continues to grow rapidly. Reference and relative dosimetry in small and composite fields are the subject of the International Atomic Energy Agency (IAEA) dosimetry code of practice that has been published as TRS-483 and an AAPM summary publication (IAEA TRS 483; Dosimetry of small static fields used in external beam radiotherapy: An IAEA/AAPM International Code of Practice for reference and relative dose determination, Technical Report Series No. 483; Pal-mans et al., Med Phys 45(11):e1123, 2018). The charge of AAPM task group 155 (TG-155) is to summarize current knowledge on small-field dosimetry and to provide recommendations of best practices for relative dose determination in small megavoltage photon beams. An overview of the issue of LCPE and the changes in photon beam perturbations with decreasing field size is provided. Recommendations are included on appropriate detector systems and measurement methodologies. Existing published data on dosimetric parameters in small photon fields (e.g., percentage depth dose, tissue phantom ratio/tissue maximum ratio, off-axis ratios, and field output factors) together with the necessary perturbation corrections for various detectors are reviewed. A discussion on errors and an uncertainty analysis in measurements is provided. The design of beam models in treatment planning systems to simulate small fields necessitates special attention on the influence of the primary beam source and collimating devices in the computation of energy fluence and dose. The general requirements for fluence and dose calculation engines suitable for modeling dose in small fields are reviewed. Implementations in commercial treatment planning systems vary widely, and the aims of this report are to provide insight for the medical physicist and guidance to developers of beams models for radiotherapy treatment planning systems.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

detector correction factors
non-equilibrium dosimetry
photon dose modeling
relative dosimetry
small field
treatment modalities

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy