Determining primary stability for adhesively stabilized dental implants
-
- Andersen, Ole Zoffmann (författare)
- Univ Bern, Dept Periodontol, Frieburg str 7, CH-3010 Bern, Switzerland.;Inst Straumann AG, Basel, Switzerland.
-
- Bellon, Benjamin (författare)
- Inst Straumann AG, Basel, Switzerland.;Univ Tampere, Fac Med & Hlth Technol, Tampere, Finland.
-
- Lamkaouchi, Maryam (författare)
- Univ Bordeaux, INSERM BIOTIS, U1026, Bordeaux, France.
-
visa fler...
-
- Brunelli, Marzia (författare)
- Inst Straumann AG, Basel, Switzerland.
-
- Wei, Qiuju (författare)
- Univ Bern, Dept Periodontol, Frieburg str 7, CH-3010 Bern, Switzerland.
-
- Procter, Philip (författare)
- Uppsala universitet,Tillämpad materialvetenskap
-
- Pippenger, Benjamin E. (författare)
- Univ Bern, Dept Periodontol, Frieburg str 7, CH-3010 Bern, Switzerland.;Inst Straumann AG, Basel, Switzerland.
-
visa färre...
-
Univ Bern, Dept Periodontol, Frieburg str 7, CH-3010 Bern, Switzerland;Inst Straumann AG, Basel, Switzerland. Inst Straumann AG, Basel, Switzerland.;Univ Tampere, Fac Med & Hlth Technol, Tampere, Finland. (creator_code:org_t)
- Springer, 2023
- 2023
- Engelska.
-
Ingår i: Clinical Oral Investigations. - : Springer. - 1432-6981 .- 1436-3771. ; 27, s. 3741-3748
- Relaterad länk:
-
https://doi.org/10.1...
-
visa fler...
-
https://uu.diva-port... (primary) (Raw object)
-
https://urn.kb.se/re...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Objectives: To examine factors influencing the primary stability of dental implants when stabilized in over-sized osteotomies using a calcium phosphate-based adhesive cement was the objective.Methods: Using implant removal torque measurements as a surrogate for primary stability, we examined the influence of implant design features (diameter, surface area, and thread design), along with cement gap size and curing time, on the resulting primary implant stability.Results: Removal torque values scaled with implant surface area and increasing implant diameters. Cement gap size did not alter the median removal torque values; however, larger gaps were associated with an increased spread of the measured values. Among the removal torque values measured, all were found to be above 32 Ncm which is an insertion torque threshold value commonly recommended for immediate loading protocols.Conclusion: The adhesive cement show potential for offering primary implant stability for different dental implant designs. In this study, the primary parameters influencing the measured removal torque values were the implant surface area and diameter. As the liquid cement prevents the use of insertion torque, considering the relationship between insertion and removal torque, removal torque can be considered a reliable surrogate for primary implant stability for bench and pre-clinical settings.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Odontologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Dentistry (hsv//eng)
Nyckelord
- Bioadhesive
- Dental implant
- Primary stability
- Removal torque
- In vitro