Sökning: WFRF:(Biffi Mauro) >
Performance evaluat...
Performance evaluation of a right atrial automatic capture verification algorithm using two different sensing configurations.
-
- Sperzel, Johannes (författare)
- Gothenburg University,Göteborgs universitet,Externa relationer,External relations
-
- Goetze, Stephan (författare)
- Gothenburg University,Göteborgs universitet,Externa relationer,External relations
-
- Kennergren, Charles, 1948 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för medicin, avdelningen för molekylär och klinisk medicin,Institute of Medicine, Department of Molecular and Clinical Medicine
-
visa fler...
-
- Biffi, Mauro (författare)
- Gothenburg University,Göteborgs universitet,Externa relationer,External relations
-
- Brooke, M Jason (författare)
- Gothenburg University,Göteborgs universitet,Externa relationer,External relations
-
- Vireca, Elisa (författare)
- Gothenburg University,Göteborgs universitet,Externa relationer,External relations
-
- Saha, Sunipa (författare)
- Gothenburg University,Göteborgs universitet,Externa relationer,External relations
-
- Schubert, Bernd (författare)
- Gothenburg University,Göteborgs universitet,Externa relationer,External relations
-
- Butter, Christian (författare)
- Gothenburg University,Göteborgs universitet,Externa relationer,External relations
-
visa färre...
-
(creator_code:org_t)
- Wiley, 2009
- 2009
- Engelska.
-
Ingår i: Pacing and clinical electrophysiology : PACE. - : Wiley. - 1540-8159 .- 0147-8389. ; 32:5, s. 579-87
- Relaterad länk:
-
https://gup.ub.gu.se...
-
visa fler...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- BACKGROUND: This acute data collection study evaluated the performance of a right atrial (RA) automatic capture verification (ACV) algorithm based on evoked response sensing from two electrode configurations during independent unipolar pacing. METHODS: RA automatic threshold tests were conducted. Evoked response signals were simultaneously recorded between the RA(Ring) electrode and an empty pacemaker housing electrode (RA(Ring)-->Can) and the electrically isolated Indifferent header electrode (RA(Ring)-->Ind). The atrial evoked response (AER) and the performance of the ACV algorithm were evaluated off-line using each sensing configuration. An accurate threshold measurement was defined as within 0.2 V of the unipolar threshold measured manually. Threshold tests were designed to fail for small AER (< 0.35 mV) or insufficient signal-to-artifact ratio (SAR < 2). Manual threshold measurements were obtained during RA unipolar and bipolar pacing and compared across device indications. RESULTS: Data were collected from 38 patients with RA bipolar leads from four manufacturers. AER signals were analyzed from 34 patients who were indicated for a pacemaker (five), implantable cardioverter-defibrillator (11), or cardiac resynchronization therapy pacemaker (six) or defibrillator (12). The minimum AER amplitude was larger (P < 0.0001) when recorded from RA(Ring)-->Can (1.6+/-0.9 mV) than from RA(Ring)-->Ind (1.3+/-0.8 mV). The algorithm successfully measured the pacing threshold in 96.8% and 91.0% of tests for RA(Ring)-->Can and RA(Ring)-->Ind, respectively. No statistical difference between the unipolar and bipolar pacing threshold was observed. CONCLUSIONS: The RA(Ring)-->Can AER sensing configuration may provide a means of implementing an independent pacing/sensing method for ACV in the RA. RA bipolar pacing therapy based on measured RA unipolar pacing thresholds may be feasible.
Nyckelord
- Aged
- Algorithms
- Atrial Fibrillation
- diagnosis
- prevention & control
- Cardiac Pacing
- Artificial
- methods
- Diagnosis
- Computer-Assisted
- methods
- Electrocardiography
- methods
- Female
- Humans
- Male
- Pattern Recognition
- Automated
- methods
- Reproducibility of Results
- Sensitivity and Specificity
- Therapy
- Computer-Assisted
- methods
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas