SwePub
Sök i LIBRIS databas

  Utökad sökning

WFRF:(Bragdon Charles R.)
 

Sökning: WFRF:(Bragdon Charles R.) > Stable fixation of ...

Stable fixation of an osseointegated implant system for above-the-knee amputees.

Nebergall, Audrey (författare)
Bragdon, Charles R, 1959 (författare)
Antonellis, Anne (författare)
visa fler...
Kärrholm, Johan, 1951 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för ortopedi,Institute of Clinical Sciences, Department of Orthopaedics
Brånemark, Rickard, 1960 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för ortopedi,Institute of Clinical Sciences, Department of Orthopaedics
Malchau, Henrik, 1951 (författare)
visa färre...
 (creator_code:org_t)
2012-04-11
2012
Engelska.
Ingår i: Acta orthopaedica. - : Medical Journals Sweden AB. - 1745-3682 .- 1745-3674. ; 83:2, s. 121-128
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background and purpose Rehabilitation of patients with transfemoral amputations is particularly difficult due to problems in using standard socket prostheses. We wanted to assess long-term fixation of the osseointegrated implant system (OPRA) using radiostereometric analysis (RSA) and periprosthetic bone remodeling. Methods 51 patients with transfemoral amputations (55 implants) were enrolled in an RSA study. RSA and plain radiographs were scheduled at 6 months and at 1, 2, 5, 7, and 10 years after surgery. RSA films were analyzed using UmRSA software. Plain radiographs were graded for bone resorption, cancellization, cortical thinning, and trabecular streaming or buttressing in specifically defined zones around the implant. Results At 5 years, the median (SE) migration of the implant was -0.02 (0.06) mm distally. The rotational movement was 0.42 (0.32) degrees around the longitudinal axis. There was no statistically significant difference in median rotation or migration at any follow-up time. Cancellization of the cortex (plain radiographic grading) appeared in at least 1 zone in over half of the patients at 2 years. However, the prevalence of cancellization had decreased by the 5-year follow-up. Interpretation The RSA analysis for the OPRA system indicated stable fixation of the implant. The periprosthetic bone remodeling showed similarities with changes seen around uncemented hip stems. The OPRA system is a new and promising approach for addressing the challenges faced by patients with transfemoral amputations.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Kirurgi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Surgery (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy