Platelet activation and degradation in an experimental extracorporeal system. A comparison between a silicone membrane and a hollow-fibre oxygenator.
-
- Mellgren, Karin, 1962 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kvinnors och barns hälsa, Avdelningen för pediatrik,Institute for the Health of Women and Children, Dept of Paediatrics
-
- Skogby, Maria, 1965 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kvinnors och barns hälsa, Avdelningen för pediatrik,Institute for the Health of Women and Children, Dept of Paediatrics
-
Järnås, A (författare)
-
visa fler...
-
Friberg, L G (författare)
-
- Wadenvik, Hans, 1955 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för invärtesmedicin, Avdelningen för internmedicin,Institute of Internal Medicine, Dept of Medicine
-
Mellgren, G (författare)
-
visa färre...
-
(creator_code:org_t)
- 1996
- 1996
- Engelska.
-
Ingår i: Perfusion. - 0267-6591. ; 11:5, s. 383-8
- Relaterad länk:
-
https://gup.ub.gu.se...
Abstract
Ämnesord
Stäng
- Blood platelets are rapidly activated in contact with biomaterials and, therefore, can be used as markers of the biocompatibility of various components in an extracorporeal system. In the present work two different oxygenators, one membrane oxygenator (Avecor) and one hollow-fibre oxygenator ("Lilliput', Dideco) were compared. Complete in vitro extracorporeal membrane oxygenation circuits were perfused with fresh, heparinized human blood for 24 h. Eight experiments were performed with the hollow-fibre oxygenator and five experiments with the membrane oxygenator. Blood gases, electrolytes, glucose and haematocrit were kept within physiological limits. Platelet count, plasma concentration of beta-thromboglobulin, platelet serotonin content, platelet membrane glycoprotein lb and its degradation product glycocalicin, as well as plasma haemoglobin concentration were assayed. As regards most of these variables, significant differences in favour of the hollow-fibre oxygenator were observed.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine (hsv//eng)
Nyckelord
- Extracorporeal Membrane Oxygenation
- instrumentation
- Humans
- Membranes
- Artificial
- Platelet Activation
- Polyvinyl Chloride
- Silicones