Evaluation of bone tissue formation in a flat surface attachment of a Bone Conduction Implant - A pilot study in a sheep model

Eeg-Olofsson, Måns, 1967 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för öron-, näs- och halssjukdomar,Institute of Clinical Sciences, Department of Otorhinolaryngology,University of Gothenburg
Lith, Agneta, 1954 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för odontologi, sektion 1,Institute of Odontology, Section 1,University of Gothenburg
Håkansson, Bo, 1953 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för odontologi, sektion 2,Institute of Odontology, Section 2
visa fler...
Reinfeldt, Sabine, 1978 (författare)
Chalmers tekniska högskola,Chalmers University of Technology
Taghavi, Hamidreza, 1981 (författare)
Chalmers tekniska högskola,Chalmers University of Technology
Fredén Jansson, Karl-Johan, 1988 (författare)
Chalmers tekniska högskola,Chalmers University of Technology
Johansson, Carina B., 1955 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för odontologi, sektion 2,Institute of Odontology, Section 2
visa färre...
 (creator_code:org_t)
2014-12-24
2014
Engelska.
Ingår i: Audiology & Neurotology Extra. - : S. Karger AG. - 1664-5537. ; 4:3, s. 62-76
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The Bone Conduction Implant (BCI) is a new bone conduction hearing device implanted under intact skin. The transducer has a flat direct contact to the mastoid part of the temporal bone and no screws are used. The sound signal is transmitted from the external audio processor to the implant by means of magnetic induction. In this study, osseointegration of a flat passive BCI transducer dummy in sheep skulls was assessed using quantitative and qualitative histology as well as Cone Beam Computed Tomography (CBCT) and Computed Tomography (CT). The histology results were also related to the mechanical properties of the bone to implant interface. Eight months after the surgical implantation, histology sections of the bone close to the implant showed bone remodelling, compact bone and some degree of osseointegration. The histological findings corresponded well to the mechanical measurements indicating stiffer bone close to the implant, and unaffected skull vibration transmission. Neither CBCT nor CT had enough resolution to visualize the bone to implant interface in detail. In this study, using an animal model, it is shown that a flat implant in contact with bone, can be a feasible method for efficient vibration transmission to the skull bone.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Medicinska och farmaceutiska grundvetenskaper -- Cell- och molekylärbiologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Basic Medicine -- Cell and Molecular Biology (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Medicinsk bioteknologi -- Biomaterialvetenskap (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Medical Biotechnology -- Biomaterials Science (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Oto-rhino-laryngologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Otorhinolaryngology (hsv//eng)

Nyckelord

Bone Conduction
Implantable bone conduction hearing devices
Bone Conduction Implant
Osseointegration
Histomorphometry
Cone Beam Computed Tomography
Bone Conduction

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy