Dose dependence of mass and microcalcification detection in digital mammography: free response human observer studies.
-
- Ruschin, Mark (author)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
-
- Timberg, Pontus (author)
- Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Medicinsk strålningsfysik, Malmö,Radiology Diagnostics, Malmö,Lund University Research Groups,Medical Radiation Physics, Malmö
-
- Båth, Magnus, 1974 (author)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
-
show more...
-
- Hemdal, Bengt (author)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
-
- Svahn, Tony (author)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
-
Saunders, Rob S (author)
-
Samei, Ehsan (author)
-
- Andersson, Ingvar (author)
- Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Radiology Diagnostics, Malmö,Lund University Research Groups
-
- Mattsson, Sören (author)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
-
Chakrabort, Dev P (author)
-
- Tingberg, Anders (author)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
-
show less...
-
(creator_code:org_t)
- 2007-01-08
- 2007
- English.
-
In: Medical physics. - : Wiley. - 0094-2405. ; 34:2, s. 400-7
- Related links:
-
https://europepmc.or...
-
show more...
-
http://www.pubmedcen... (free)
-
http://dx.doi.org/10...
-
https://gup.ub.gu.se...
-
https://doi.org/10.1...
-
https://lup.lub.lu.s...
-
show less...
Abstract
Subject headings
Close
- The purpose of this study was to evaluate the effect of dose reduction in digital mammography on the detection of two lesion types-malignant masses and clusters of microcalcifications. Two free-response observer studies were performed-one for each lesion type. Ninety screening images were retrospectively selected; each image was originally acquired under automatic exposure conditions, corresponding to an average glandular dose of 1.3 mGy for a standard breast (50 mm compressed breast thickness with 50% glandularity). For each study, one to three simulated lesions were added to each of 40 images (abnormals) while 50 were kept without lesions (normals). Two levels of simulated system noise were added to the images yielding two new image sets, corresponding to simulated dose levels of 50% and 30% of the original images (100%). The manufacturer's standard display processing was subsequently applied to all images. Four radiologists experienced in mammography evaluated the images by searching for lesions and marking and assigning confidence levels to suspicious regions. The search data were analyzed using jackknife free-response (JA-FROC) methodology. For the detection of masses, the mean figure-of-merit (FOM) averaged over all readers was 0.74, 0.71, and 0.68 corresponding to dose levels of 100%, 50%, and 30%, respectively. These values were not statistically different from each other (F= 1.67, p=0.19) but showed a decreasing trend. In contrast, in the microcalcification study the mean FOM was 0.93, 0.67, and 0.38 for the same dose levels and these values were all significantly different from each other (F = 109.84, p < 0.0001). The results indicate that lowering the present dose level by a factor of two compromised the detection of microcalcifications but had a weaker effect on mass detection.
Subject headings
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
- NATURVETENSKAP -- Biologi -- Biofysik (hsv//swe)
- NATURAL SCIENCES -- Biological Sciences -- Biophysics (hsv//eng)
Keyword
- Breast Neoplasms
- radiography
- Calcinosis
- radiography
- Dose-Response Relationship
- Radiation
- Image Enhancement
- methods
- Image Interpretation
- Computer-Assisted
- methods
- Mammography
- methods
- Observer Variation
- Precancerous Conditions
- radiography
- Radiation Dosage
- Radiographic Image Enhancement
- methods
- Reproducibility of Results
- Sensitivity and Specificity
- observer performance
- free response
- dose reduction
- Digital mammography
Publication and Content Type
- ref (subject category)
- art (subject category)
Find in a library
To the university's database
- By the author/editor
-
Ruschin, Mark
-
Timberg, Pontus
-
Båth, Magnus, 19 ...
-
Hemdal, Bengt
-
Svahn, Tony
-
Saunders, Rob S
-
show more...
-
Samei, Ehsan
-
Andersson, Ingva ...
-
Mattsson, Sören
-
Chakrabort, Dev ...
-
Tingberg, Anders
-
show less...
- About the subject
-
- MEDICAL AND HEALTH SCIENCES
-
MEDICAL AND HEAL ...
-
and Clinical Medicin ...
-
and Radiology Nuclea ...
-
- NATURAL SCIENCES
-
NATURAL SCIENCES
-
and Biological Scien ...
-
and Biophysics
- Articles in the publication
-
Medical physics
- By the university
-
University of Gothenburg
-
Lund University