Improved in-plane visibility of tumors using breast tomosynthesis

Ruschin, Mark (author)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Timberg, Pontus (author)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Svahn, Tony (author)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
show more...
Andersson, Ingvar (author)
Lund University,Lunds universitet,Institutionen för translationell medicin,Medicinska fakulteten,Department of Translational Medicine,Faculty of Medicine
Hemdal, Bengt (author)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Mattsson, Sören (author)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Båth, Magnus, 1974 (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Tingberg, Anders (author)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
show less...
 (creator_code:org_t)
SPIE, 2007
2007
English.
In: Progress in Biomedical Optics and Imaging - Proceedings of SPIE. - : SPIE. - 1605-7422 .- 1042-4687. ; 6510
  • Journal article (other academic/artistic)
Abstract Subject headings
Close  
  • The purpose of this work was to evaluate and compare the visibility of simulated tumors in 2D digital mammography (DM) and breast tomosynthesis (BT) images of patients. Images of the same women were acquired on both a DM system (Mammomat Novation, Siemens) and a BT prototype system adapted from the same type of DM system. Using the geometrical properties of the two systems, simulated, ellipsoid-shaped tumors (average dimension: 8.4 mm × 6.6 mm × 5 mm) with irregular margins were projected and added to each DM image as well as each BT projection image prior to 3D reconstruction. The same beam quality and approximately the same total absorbed dose were used for each breast image acquisition on both systems. Two simulated tumors were added to each of thirty patient scans, yielding sixty cases. A series of 4-alternative forced choice (4-AFC) human observer experiments were conducted in order to determine what projected signal intensity (contrast) of the tumors in the DM images would be needed to achieve the same detectability as in the reconstructed BT images. Nine observers participated. For the BT 4-AFC experiment, when the signal intensity of the tumor on the central projection was 0.010 (natural logarithmic units) the mean percent of correct responses (PC) was measured to be 81.5%, which converted to a detectability index value (d') of 1.96. For the DM system, the same detectability was achieved at a signal intensity determined to be 0.038. Equivalent levels of tumor detection in BT images were thus achieved at around four times less projected signal intensity than in DM images, indicating that the use of BT may lead to earlier detection of breast cancer.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Keyword

Alternative forced choice
Detection
Digital breast tomosynthesis
Digital mammography
Signal intensity

Publication and Content Type

vet (subject category)
art (subject category)

Find in a library

To the university's database

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view