Insufficient mixing of thawed serum samples leading to erroneous results - experience from a field study and use of a correction procedure
-
- Ekström, Ulf (författare)
- Lund University,Lunds universitet,Avdelningen för klinisk kemi och farmakologi,Institutionen för laboratoriemedicin,Medicinska fakulteten,Division of Clinical Chemistry and Pharmacology,Department of Laboratory Medicine,Faculty of Medicine,La Isla Network
-
- Apelqvist, Jenny (författare)
- Lund University,Lunds universitet,Avdelningen för klinisk kemi och farmakologi,Institutionen för laboratoriemedicin,Medicinska fakulteten,Division of Clinical Chemistry and Pharmacology,Department of Laboratory Medicine,Faculty of Medicine
-
- Hansson, Erik, 1987 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för medicin, avdelningen för samhällsmedicin och folkhälsa,Institute of Medicine, School of Public Health and Community Medicine,Sahlgrenska Academy
-
visa fler...
-
- Bodin, Theo (författare)
- Karolinska Institutet,Karolinska Institute
-
- Wegman, D. H. (författare)
- University of Massachusetts Lowell
-
- Abrahamson, Magnus (författare)
- Lund University,Lunds universitet,Avdelningen för klinisk kemi och farmakologi,Institutionen för laboratoriemedicin,Medicinska fakulteten,Division of Clinical Chemistry and Pharmacology,Department of Laboratory Medicine,Faculty of Medicine
-
- Jakobsson, K. (författare)
- Sahlgrenska University Hospital
-
visa färre...
-
(creator_code:org_t)
- 2019-12-18
- 2020
- Engelska.
-
Ingår i: Scandinavian Journal of Clinical & Laboratory Investigation. - : Informa UK Limited. - 0036-5513 .- 1502-7686. ; 80:2, s. 99-105
- Relaterad länk:
-
http://dx.doi.org/10...
-
visa fler...
-
https://gup.ub.gu.se...
-
https://doi.org/10.1...
-
https://lup.lub.lu.s...
-
http://kipublication...
-
visa färre...
Abstract
Ämnesord
Stäng
- Incorrect analysis results that are close to expected might not be recognized in scientific studies or routine patient care. In two field studies we obtained unexpected results in a large number of samples. The present study aimed to identify the source of error in the samples from these studies and to validate a method to obtain correct results. Pre-analytical procedures were scrutinized, giving no indications of inappropriate pre-analytical sample handling in the field or during transport in a tropical climate. Using a new set of samples from volunteers in simulation experiments, we observed the known concentration gradient of analytes sampled in gel as well as plain tubes after freezer storage and thawing. Experiments demonstrated that mixing of samples by vortexing alone was not sufficient to disrupt the gradient formed by freezing and thawing, which appeared to cause the problem encountered when we in field studies analyzed and biobanked large sample sets by robot pipetting. A correction procedure was introduced, in which the obtained value of an analyte was multiplied by a correction factor calculated for each sample using the expected sodium level (140 mmol/L) divided by the measured sodium value. When it was validated on results from the simulation experiments, we repeatedly found that the correction lead to results very close to true values for analytes of different size and charge. Usefulness of the procedure was demonstrated when applied to a large set of field study results.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine (hsv//eng)
- NATURVETENSKAP -- Kemi -- Analytisk kemi (hsv//swe)
- NATURAL SCIENCES -- Chemical Sciences -- Analytical Chemistry (hsv//eng)
Nyckelord
- Chemistry
- clinical
- clinical laboratory techniques
- diagnostic tests
- routine
- Pre-analytical phase
- analytic sample preparation methods
- specimen handling
- intervention
- workers
- Research & Experimental Medicine