SwePub
Sök i LIBRIS databas

  Utökad sökning

L773:0147 8389
 

Sökning: L773:0147 8389 > Mechanical extracti...

Mechanical extraction of cardiac implantable electronic devices leads with long dwell time: Efficacy and safety of the step up approach

Lensvelt, L. M. H. (författare)
Egorova, A. D. (författare)
Schalij, M. J. (författare)
visa fler...
Yilmaz, D. (författare)
Kennergren, Charles, 1948 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för medicin, avdelningen för molekylär och klinisk medicin,Institute of Medicine, Department of Molecular and Clinical Medicine
Bootsma, M. (författare)
van Erven, L. (författare)
visa färre...
 (creator_code:org_t)
2020-11-12
2021
Engelska.
Ingår i: Pacing and Clinical Electrophysiology: PACE. - : Wiley. - 0147-8389 .- 1540-8159. ; 44:1, s. 120-128
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The aim of this study was to evaluate the efficacy and safety of the stepwise mechanical transvenous lead extraction approach in a patient population with chronically implanted transvenous leads with a long dwell time. From January 2014 till December 2018, all lead extractions with lead dwell time >= 5 years performed at our tertiary centre were retrospectively analysed. A total of 173 leads, from 78 patients (median age 68 years; 81% male) with a median dwell time of 9 years (interquartile range [IQR] 5) were extracted, with three or more leads in 42% of the patients. Right atrial leads: 41%; right ventricular pacing leads: 16%; implantable cardioverter-defibrillator (ICD) leads: 31% (72% dual coil); coronary sinus leads: 12%. The majority (75%) of the leads had an active fixation. Most frequent indication for extraction was pocket infection/erosion (76%). Overall clinical success was 97%, and complete procedural success was 93%. Venous patency, assessed with venous angiography, was well preserved in 93% of the cases. The overall procedural complication rate was 3.8% (2.6% major and 1.3% minor). Despite the complexity of the population and a very long dwell time (median 9 years), a clinical success rate of 97% was achieved with the stepwise mechanical approach. Analysis of impeding progression of pectoral extraction suggests that dense fibrosis and sharp lead curvature in the transvenous trajectory pose a challenge. Complication rate was low, and acute venous patency was generally well preserved.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine (hsv//eng)

Nyckelord

cardiac implantable devices
device-related complications
extraction
tool
mechanical sheath
TightRail
transvenous lead extractions
prevalence
Cardiovascular System & Cardiology
Engineering

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)