Synthetic computed ...
Synthetic computed tomography data allows for accurate absorbed dose calculations in a magnetic resonance imaging only workflow for head and neck radiotherapy
-
- Palmér, Emilia, 1993 (författare)
- University of Gothenburg,Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för medicinsk strålningsvetenskap,Institute of Clinical Sciences, Department of Medical Radiation Sciences
-
- Karlsson, Anna (författare)
- University of Gothenburg,Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för medicinsk strålningsvetenskap,Institute of Clinical Sciences, Department of Medical Radiation Sciences,Sahlgrenska University Hospital
-
- Nordström, Fredrik, 1981 (författare)
- University of Gothenburg,Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för medicinsk strålningsvetenskap,Institute of Clinical Sciences, Department of Medical Radiation Sciences,Sahlgrenska University Hospital
-
visa fler...
-
- Petruson, Karin, 1967 (författare)
- University of Gothenburg,Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för onkologi,Institute of Clinical Sciences, Department of Oncology
-
- Siversson, Carl (författare)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups,Spectronic Medical AB
-
- Ljungberg, Maria (författare)
- University of Gothenburg,Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för medicinsk strålningsvetenskap,Institute of Clinical Sciences, Department of Medical Radiation Sciences,Sahlgrenska University Hospital
-
- Sohlin, Maja, 1982 (författare)
- University of Gothenburg,Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för medicinsk strålningsvetenskap,Institute of Clinical Sciences, Department of Medical Radiation Sciences,Sahlgrenska University Hospital
-
visa färre...
-
(creator_code:org_t)
- Elsevier BV, 2021
- 2021
- Engelska.
-
Ingår i: Physics and Imaging in Radiation Oncology. - : Elsevier BV. - 2405-6316. ; 17, s. 36-42
- Relaterad länk:
-
http://phiro.science...
-
visa fler...
-
http://dx.doi.org/10... (free)
-
https://gup.ub.gu.se...
-
https://doi.org/10.1...
-
https://lup.lub.lu.s...
-
visa färre...
Abstract
Ämnesord
Stäng
- Background and purpose: Few studies on magnetic resonance imaging (MRI) only head and neck radiation treatment planning exist, and none using a generally available software. The aim of this study was to evaluate the accuracy of absorbed dose for head and neck synthetic computed tomography data (sCT) generated by a commercial convolutional neural network-based algorithm. Materials and methods: For 44 head and neck cancer patients, sCT were generated and the geometry was validated against computed tomography data (CT). The clinical CT based treatment plan was transferred to the sCT and recalculated without re-optimization, and differences in relative absorbed dose were determined for dose-volume-histogram (DVH) parameters and the 3D volume. Results: For overall body, the results of the geometric validation were (Mean ± 1sd): Mean error −5 ± 10 HU, mean absolute error 67 ± 14 HU, Dice similarity coefficient 0.98 ± 0.05, and Hausdorff distance difference 4.2 ± 1.7 mm. Water equivalent depth difference for region Th1-C7, mid mandible and mid nose were −0.3 ± 3.4, 1.1 ± 2.0 and 0.7 ± 3.8 mm respectively. The maximum mean deviation in absorbed dose for all DVH parameters was 0.30% (0.12 Gy). The absorbed doses were considered equivalent (p-value < 0.001) and the mean 3D gamma passing rate was 99.4 (range: 95.7–99.9%). Conclusions: The convolutional neural network-based algorithm generates sCT which allows for accurate absorbed dose calculations for MRI-only head and neck radiation treatment planning. The sCT allows for statistically equivalent absorbed dose calculations compared to CT based radiotherapy.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Cancer och onkologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Cancer and Oncology (hsv//eng)
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Nyckelord
- Convolutional neural network
- Head and neck
- MRI only
- Pseudo CT
- Synthetic CT
- Treatment planning
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas
- Av författaren/redakt...
-
Palmér, Emilia, ...
-
Karlsson, Anna
-
Nordström, Fredr ...
-
Petruson, Karin, ...
-
Siversson, Carl
-
Ljungberg, Maria
-
visa fler...
-
Sohlin, Maja, 19 ...
-
visa färre...
- Om ämnet
-
- MEDICIN OCH HÄLSOVETENSKAP
-
MEDICIN OCH HÄLS ...
-
och Klinisk medicin
-
och Cancer och onkol ...
-
- MEDICIN OCH HÄLSOVETENSKAP
-
MEDICIN OCH HÄLS ...
-
och Klinisk medicin
-
och Radiologi och bi ...
- Artiklar i publikationen
-
Physics and Imag ...
- Av lärosätet
-
Göteborgs universitet
-
Lunds universitet