Comparison of color LCD and medical-grade monochrome LCD displays in diagnostic radiology.
-
Geijer, H (författare)
-
- Geijer, Mats, 1957 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper,Institute of Clinical Sciences
-
- Forsberg, Lillemor (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper,Institute of Clinical Sciences
-
visa fler...
-
- Kheddache, Susanne, 1949 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper,Institute of Clinical Sciences
-
Sund, P (författare)
-
visa färre...
-
(creator_code:org_t)
- 2007-03-06
- 2007
- Engelska.
-
Ingår i: Journal of Digital Imaging. - : Springer Science and Business Media LLC. - 0897-1889 .- 1618-727X. ; 20:2, s. 114-121
- Relaterad länk:
-
https://europepmc.or...
-
visa fler...
-
https://gup.ub.gu.se...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Abstract In diagnostic radiology, medical-grade monochrome displays are usually recommended because of their higher luminance. Standard color displays can be used as a less expensive alternative, but have a lower luminance. The aim of the present study was to compare image quality for these two types of displays. Images of a CDRAD contrast-detail phantom were read by four radiologists using a 2-megapixel (MP) color display (143 cd/m2 maximum luminance) as well as 2-MP (295 cd/m2) and 3-MP monochrome displays. Thirty lumbar spine radiographs were also read by four radiologists using the color and the 2-MP monochrome display in a visual grading analysis (VGA). Very small differences were found between the displays when reading the CDRAD images. The VGA scores were −0.28 for the color and −0.25 for the monochrome display (p=0.24; NS). It thus seems possible to use color displays in diagnostic radiology provided that grayscale adjustment is used. Key words PACS - displays - digital imaging - luminance - image quality - monitor - medical imaging - liquid crystal display
Nyckelord
- PACS
- displays
- digital imaging
- luminance
- image quality
- monitor
- medical imaging
- liquid crystal display