SwePub
Sök i LIBRIS databas

  Utökad sökning

WFRF:(Alian F)
 

Sökning: WFRF:(Alian F) > Linearity and Lever...

Linearity and Lever Ratio of the Normal and Reconstructed Cadaveric Human Middle Ear

Majdalawieh, Osama F (författare)
Alian, Wael A, 1970 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper,Institute of Clinical Sciences
Katlai, Balaji (författare)
visa fler...
Van Wijhe, Rene (författare)
Bance, Manohar (författare)
visa färre...
 (creator_code:org_t)
2008
2008
Engelska.
Ingår i: Otology & Neurotology. - 1531-7129. ; 29:6, s. 796-802
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Hypothesis: The linearity and the level ratio are different in reconstructed ears. Background: The linearity of the reconstructed human middle ear (ME) has not been previously explored. It is important to analyze if high sound pressure levels (SPLs) result in distortion due to nonlinearities particularly because hearing aids have high-output SPLs. The diseased ME is reconstructed with prostheses. These diseased ears frequently need additional amplification with hearing aids, and it is unclear if reconstruction itself leads to nonlinear ME responses. Methods: Eight fresh human cadaveric temporal bones were used. Pure tones of 70, 90, and 110 dB SPL at 500, 1,000, and 3,000 Hz were presented to the ear canal. Umbo and stapes displacements were measured by means of a laser Doppler vibrometer. After removing the incus, the tympanic membrane assembly to the stapes head prosthesis was placed, and measurements were repeated. Results: Stapes footplate vibrations in the reconstructed ears are 10 to 15 dB lower than those of the normal ears. In both normal and reconstructed ears, the footplate vibrations are linearly related to SPL at the tympanic membrane between 70 and 110 dB SPL at the frequencies tested. For the lever ratio, intact ears are more efficient at transmission of umbo vibrations to the stapes compared with reconstructed ears. Conclusion: To within acceptable limits, the ME seems to be linear between 70 and 110 dB SPL input levels, across the speech frequencies, and this does not change with reconstruction. The reconstructed human ME seems to have a less efficient lever ratio than the intact ME.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Oto-rhino-laryngologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Otorhinolaryngology (hsv//eng)

Nyckelord

laser doppler vibrometry
lever ratio
linear systems
middle ear mechanics
ossiculoplasty
partial ossicular replacement prosthesis

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy