The influence of the characteristic curve on the clinical image quality and patient absorbed dose in lumbar spine radiography

Tingberg, Anders (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Herrmann, C. (författare)
Lanhede, B. (författare)
visa fler...
Almén, A (författare)
Mattsson, Sören (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Panzer, W. (författare)
Besjakov, Jack (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Radiology Diagnostics, Malmö,Lund University Research Groups
Månsson, L.G. (författare)
Kheddache, S. (författare)
Zankl, M. (författare)
visa färre...
 (creator_code:org_t)
SPIE, 2001
2001
Engelska.
Ingår i: Proceedings of SPIE. - : SPIE. - 0819440108 ; 4324, s. 41503-41503
  • Konferensbidrag (refereegranskat)
Abstract Ämnesord
Stäng  
  • The 'European Guidelines on Quality Criteria for Diagnostic Radiographic Images' do not address the choice of film characteristic (H/D) curve, which is an important parameter for the description of a radiographic screen-film system. Since it is not possible to investigate this influence by taking repeated exposures of the same patients on films with systematically varied H/D curves, patient images of lumbar spine were digitised in the current study. The image contrast was altered by digital image processing techniques, simulating images with H/D curves varying from flat over standard latitude to a film type steeper than a mammography film. The manipulated images were printed on film for evaluation. Seven European radiologists evaluated the clinical image quality of in total 224 images by analysing the fulfilment of the European Image Criteria and by visual grading analysis of the images. The results show that the local quality can be significantly improved by the application of films with a steeper film H/D curve compared to the standard latitude film. For images with an average optical density of about 1.25, the application of the steeper film results in a reduction of patient absorbed dose by about 10-15% without a loss of diagnostically relevant image information. The results also show that the patient absorbed dose reduction obtained by altering the tube voltage from 70 kV to 90 kV coincides with a loss of image information that cannot be compensated for by simply changing the shape of the H/D curve.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

kon (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy