Evaluation of 3D printed contrast detail phantoms for mammography quality assurance

Boll, Måns (författare)
Lund University,Skåne University Hospital
Vent, Trevor (författare)
Hospital of the University of Pennsylvania
Tomic, Hanna (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Medicinsk strålningsfysik, Malmö,LUCC: Lunds universitets cancercentrum,Övriga starka forskningsmiljöer,Radiology Diagnostics, Malmö,Lund University Research Groups,Medical Radiation Physics, Malmö,LUCC: Lund University Cancer Centre,Other Strong Research Environments,Skåne University Hospital
visa fler...
Bernhardsson, Christian (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups,Skåne University Hospital
Dustler, Magnus (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Medicinsk strålningsfysik, Malmö,LUCC: Lunds universitets cancercentrum,Övriga starka forskningsmiljöer,Radiology Diagnostics, Malmö,Lund University Research Groups,Medical Radiation Physics, Malmö,LUCC: Lund University Cancer Centre,Other Strong Research Environments,Skåne University Hospital
Tingberg, Anders (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,LUCC: Lunds universitets cancercentrum,Övriga starka forskningsmiljöer,Medical Radiation Physics, Malmö,Lund University Research Groups,LUCC: Lund University Cancer Centre,Other Strong Research Environments,Skåne University Hospital
Bakic, Predrag R. (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,LUCC: Lunds universitets cancercentrum,Övriga starka forskningsmiljöer,Radiology Diagnostics, Malmö,Lund University Research Groups,LUCC: Lund University Cancer Centre,Other Strong Research Environments,Hospital of the University of Pennsylvania,Skåne University Hospital
Bosmans, Hilde (redaktör/utgivare)
Marshall, Nicholas (redaktör/utgivare)
Van Ongeval, Chantal (redaktör/utgivare)
visa färre...
 (creator_code:org_t)
SPIE, 2022
2022
Engelska.
Ingår i: 16th International Workshop on Breast Imaging, IWBI 2022. - : SPIE. - 0277-786X .- 1996-756X. - 9781510655843 ; 12286
  • Konferensbidrag (refereegranskat)
Abstract Ämnesord
Stäng  
  • Objects created by 3D printers are increasingly used in various medical applications. Today, affordable 3D printers, using Fused Deposition Modeling are widely available. In this project, a commercially available 3D printer was used to replicate a conventional radiographic contrast detail phantom. Printing materials were selected by comparing their x-ray attenuation properties. Two replicas were printed using polylactic acid, with different filling patterns. The printed phantoms were imaged by a clinical mammography system, using automatic exposure control. Phantom images were visually and quantitively compared to images of the corresponding conventional contrast detail phantom. Visual scoring of the contrast detail elements was performed by a medical physics student. Contrast-to-noise ratio (CNR) was calculated for each phantom element. The diameter and thickness of the smallest visible phantom object were 0.44 mm and 0.09 mm, respectively, for both filling patterns. For the conventional phantom, the diameter and thickness of the smallest visible object were 0.31 mm and 0.09 mm. Visual inspection of printed phantoms revealed some linear artefacts. These artefacts were however not visible on mammographic projections. Quantitively, average CNR of printed phantom objects followed the same trend with an increase of average CNR with increasing disk height. However, there is a limitation of detail objects with disk diameters below 1.25 mm, caused by the available nozzle size. Based upon the encouraging results, future work will explore the use of different materials and smaller nozzle diameters.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Image Processing (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

3D printing
contrast detail phantom
contrast-tonoise ratio
Mammography quality assurance
visual analysis

Publikations- och innehållstyp

kon (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy