Denoising of arterial spin labeling data: wavelet-domain filtering compared with Gaussian smoothing.
-
- Bibic, Adnan (author)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
- Knutsson, Linda (author)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,MR Physics,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Lund University Research Groups
-
- Ståhlberg, Freddy (author)
- Lund University,Lunds universitet,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medicinsk strålningsfysik, Lund,Institutionen för kliniska vetenskaper, Lund,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Medical Radiation Physics, Lund,Department of Clinical Sciences, Lund
-
show more...
-
- Wirestam, Ronnie (author)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
show less...
-
(creator_code:org_t)
- 2010-04-28
- 2010
- English.
-
In: Magma. - : Springer Science and Business Media LLC. - 1352-8661. ; 23:3, s. 125-137
- Related links:
-
https://portal.resea... (primary) (free)
-
show more...
-
http://www.ncbi.nlm....
-
http://dx.doi.org/10...
-
https://lup.lub.lu.s...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
show less...
Abstract
Subject headings
Close
- PURPOSE: To investigate a wavelet-based filtering scheme for denoising of arterial spin labeling (ASL) data, potentially enabling reduction of the required number of averages and the acquisition time. METHODS: ASL magnetic resonance imaging (MRI) provides quantitative perfusion maps by using arterial water as an endogenous tracer. The signal difference between a labeled image, where inflowing arterial spins are inverted, and a control image is proportional to blood perfusion. ASL perfusion maps suffer from low SNR, and the experiment must be repeated a number of times (typically more than 40) to achieve adequate image quality. In this study, systematic errors introduced by the proposed wavelet-domain filtering approach were investigated in simulated and experimental image datasets and compared with conventional Gaussian smoothing. RESULTS: Application of the proposed method enabled a reduction of the number of averages and the acquisition time by at least 50% with retained standard deviation, but with effects on absolute CBF values close to borders and edges. CONCLUSIONS: When the ASL perfusion maps showed moderate-to-high SNRs, wavelet-domain filtering was superior to Gaussian smoothing in the vicinity of borders between gray and white matter, while Gaussian smoothing was a better choice for larger homogeneous areas, irrespective of SNR.
Subject headings
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Publication and Content Type
- art (subject category)
- ref (subject category)
Find in a library
-
Magma
(Search for host publication in LIBRIS)
To the university's database