Implementing safe and robust Total Marrow Irradiation using Helical Tomotherapy – A practical guide

Haraldsson, André (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Radiotherapy Physics,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Lund University Research Groups,Skåne University Hospital
Engellau, Jacob (författare)
Lund University,Lunds universitet,Tumörmikromiljö,Sektion I,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Tumor microenvironment,Section I,Department of Clinical Sciences, Lund,Faculty of Medicine,Skåne University Hospital
Lenhoff, Stig (författare)
Skåne University Hospital
visa fler...
Engelholm, Silke (författare)
Skåne University Hospital
Bäck, Sven (författare)
Skåne University Hospital
Engström, Per E. (författare)
Skåne University Hospital
visa färre...
 (creator_code:org_t)
Elsevier BV, 2019
2019
Engelska 6 s.
Ingår i: Physica Medica. - : Elsevier BV. - 1120-1797. ; 60, s. 162-167
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Total Marrow Irradiation (TMI) with Helical Tomotherapy is a radiotherapy treatment technique that targets bone marrow and sanctuary sites prior to stem cell or bone marrow transplantation (SCT/BMT). TMI is a complex procedure that involves several critical steps that all need to be carefully addressed for a successful implementation, such as dose homogeneity in field junctions, choice of target margins, integrity of treatment and back-up planning. In this work we present our solution for a robust and reproducible workflow throughout the treatment chain and data for twenty-three patients treated to date. Material & Methods: Patients were immobilized in a whole body vacuum cushion and thermoplastic mask. CT-scanning and treatment were performed in two parts with field matching at the upper thigh. Target consisted of marrow containing bone and sanctuary sites. Lungs, kidneys, bowel, heart and liver were defined as organs at risk (OAR). A fast surface scanning system was used to position parts of the body not covered by the imaging system (MVCT) as well as to reduce treatment time. Results: All patients completed their treatment and could proceed with SCT/BMT. Doses to OARs were significantly reduced and target dose homogeneity was improved compared to TBI. Robustness tests performed on field matching and patient positioning support that the field junction technique is adequate. Replacing MVCT with optical surface scanning reduced the treatment time by 25 min per fraction. Conclusion: The methodology presented here has shown to provide a safe, robust and reproducible treatment for Total Marrow Irradiation using Tomotherapy.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Helical
Irradiation
Marrow
Radiotherapy
TMI
Tomotherapy
Total

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy