A comparison of serum and EDTA plasma in the measurement of glutamic acid decarboxylase autoantibodies (GADA) and autoantibodies to islet antigen-2 (IA-2A) using the RSR radioimmunoassay (RIA) and enzyme linked immunosorbent assay (ELISA) kits.

Rahmati, Kobra (författare)
Lernmark, Åke (författare)
Lund University,Lunds universitet,Celiaki och diabetes,Forskargrupper vid Lunds universitet,Celiac Disease and Diabetes Unit,Lund University Research Groups
Becker, Charlotte (författare)
Lund University,Lunds universitet,Klinisk kemi, Malmö,Forskargrupper vid Lunds universitet,Clinical Chemistry, Malmö,Lund University Research Groups
visa fler...
Foltyn Zadura, Anna (författare)
Lund University,Lunds universitet,Klinisk kemi, Malmö,Forskargrupper vid Lunds universitet,Clinical Chemistry, Malmö,Lund University Research Groups
Larsson, Karin (författare)
Ivarsson, Sten (författare)
Lund University,Lunds universitet,Pediatrisk endokrinologi,Forskargrupper vid Lunds universitet,Paediatric Endocrinology,Lund University Research Groups
Törn, Carina (författare)
Lund University,Lunds universitet,Celiaki och diabetes,Forskargrupper vid Lunds universitet,Celiac Disease and Diabetes Unit,Lund University Research Groups
visa färre...
 (creator_code:org_t)
2008
2008
Engelska.
Ingår i: Clinical Laboratory. - 1433-6510. ; 54:7-8, s. 227-235
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • BACKGROUND: Glutamic acid decarboxylase antibodies (GADA) and tyrosine phosphatase antibodies (islet antigen-2 antibodies; IA-2A) are used in clinical practise to identify type 1 diabetes. METHODS: GADA and IA-2A were measured with RSR-ELISA kits in samples from 76 newly diagnosed type 1 diabetic children and 120 healthy controls. The aim was to evaluate performance of RSR-ELISA kits for GADA and IA-2A when serum and Ca2+ treated plasma were used. RESULTS: GADA achieved high area under the curve (AUC) both for serum 0.95 (95% CI 0.90-0.99) and for Ca2+ treated plasma 0.95 (95% CI 0.91-0.99). At specificity 98%, sensitivity was 84% for serum and 87% for Ca2+ treated plasma. IA-2A achieved AUC 0.92 (95% CI 0.87-0.97) for serum and 0.94 (95% CI 0.90-0.98) for Ca2+ treated plasma. Using the lowest standard (15 WHO-Units/ml) as cut-off, specificity for serum was 100% and for Ca2+ treated plasma 99% with sensitivity 74% in both cases. Sensitivity was higher in ELISA compared to RIA (74%; p = 0.0080) for GADA measurement and similar for ELISA and RIA IA-2A measurements (76%; p = 0.50). CONCLUSION: Both RSR-ELISAs, GADA and IA-2A showed excellent performance for serum as well as for Ca2+ treated plasma.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Endokrinologi och diabetes (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Endocrinology and Diabetes (hsv//eng)

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy