Buffer medium exchange in continuous cell and particle streams using ultrasonic standing wave focusing

Augustsson, Per (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
Åberg, Lena B (författare)
Sward-Nilsson, Ann-Margret K. (författare)
visa fler...
Laurell, Thomas (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
visa färre...
 (creator_code:org_t)
2008-07-18
2009
Engelska.
Ingår i: Microchimica Acta. - : Springer Science and Business Media LLC. - 1436-5073 .- 0026-3672. ; 164:3-4, s. 269-277
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • A microfluidic strategy to perform buffer exchange of particle and cell suspensions in a continuous flow format on, chip is presented. Ultrasonic standing wave technology is utilized to confine particulate matter to the centre of a buffer exchange channel while particle free buffer is sequentially aspirated via capillaries that branch off from the buffer exchange channel. At each such branch, clean buffer is supplied at an equal flow-rate from a capillary at the opposing channel wall, generating a sideways translation of the original buffer, laminated with a wash buffer stream. Each such junction increases the buffer exchange ratio accordingly. The reported buffer exchange system provides means to adjust buffer exchange conditions on-line by tuning the ratio of the cross-flow wash buffer relative the sample suspension flow, rate. The system performance was evaluated using 5 mu m polystyrene microbeads and a dye as the model contaminant. Wash efficiencies up to 96.4% were accomplished with a 0.2% solid content bead suspension, using eight cross-flow junctions, effectively exchanging the carrier buffer twice. The corresponding data for erythrocyte washing was recorded to be 98.3% at a haematocrit of 2%.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering (hsv//eng)

Nyckelord

Buffer medium exchange
radiation force
Acoustic
Erythrocytes
Ultrasonic standing wave
Particle washing

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Augustsson, Per
Åberg, Lena B
Sward-Nilsson, A ...
Laurell, Thomas
Om ämnet
TEKNIK OCH TEKNOLOGIER
TEKNIK OCH TEKNO ...
och Medicinteknik
Artiklar i publikationen
Microchimica Act ...
Av lärosätet
Lunds universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy