BE AWARE OF NEUTRONS OUTSIDE SHORT MAZES FROM 10-MV LINEAR ACCELERATORS X-RAYS IN RADIOTHERAPY FACILITIES.
-
- Brockstedt, Sara (författare)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
- Holstein, Hanna (författare)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
-
Jakobsson, L (författare)
-
visa fler...
-
- Tomaszewicz, Andrej (författare)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
- Knöös, Tommy (författare)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
visa färre...
-
(creator_code:org_t)
- 2015-03-22
- 2015
- Engelska.
-
Ingår i: Radiation Protection Dosimetry. - : Oxford University Press (OUP). - 1742-3406 .- 0144-8420. ; 165:1-4, s. 464-467
- Relaterad länk:
-
http://www.ncbi.nlm....
-
visa fler...
-
http://dx.doi.org/10...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- During the radiation survey of a reinstalled 10-MV linear accelerator in an old radiation treatment facility, high dose rates of neutrons were observed. The area outside the maze entrance is used as a waiting room where patients, their relatives and staff other than those involved in the actual treatment can freely pass. High fluence rates of neutrons would cause an unnecessary high effective dose to the staff working in the vicinity of such a system, and it can be several orders higher than the doses received due to X-rays at the same location. However, the common knowledge appears to have been that the effect of neutrons at 10-MV X-ray linear accelerator facilities is negligible and shielding calculations models seldom mention neutrons for this operating energy level. Although data are scarce, reports regarding this phenomenon are now emerging. For the future, it is advocated that contributions from neutrons are considered already during the planning stage of new or modified facilities aimed for 10 MV and that estimated dose levels are verified.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)