Sökning: WFRF:(Reistad N.) >
Identification of t...
Identification of tumor margins using diffuse reflectance spectroscopy with an extended-wavelength spectrum in a porcine model
-
- Dahlstrand, U (författare)
- Lund University,Lunds universitet,Oftalmologi, Lund,Sektion IV,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Forskargruppen för oftalmologisk avbildning,Forskargrupper vid Lunds universitet,Ophthalmology, Lund,Section IV,Department of Clinical Sciences, Lund,Faculty of Medicine,Ophthalmology Imaging Research Group,Lund University Research Groups,Skåne University Hospital
-
- Sheikh, R (författare)
- Lund University,Lunds universitet,Oftalmologi, Lund,Sektion IV,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Forskargruppen för oftalmologisk avbildning,Forskargrupper vid Lunds universitet,Ophthalmology, Lund,Section IV,Department of Clinical Sciences, Lund,Faculty of Medicine,Ophthalmology Imaging Research Group,Lund University Research Groups,Skåne University Hospital
-
- Nguyen, C D (författare)
- Lund University,Lunds universitet,Oftalmologi, Lund,Sektion IV,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Forskargruppen för oftalmologisk avbildning,Forskargrupper vid Lunds universitet,Ophthalmology, Lund,Section IV,Department of Clinical Sciences, Lund,Faculty of Medicine,Ophthalmology Imaging Research Group,Lund University Research Groups
-
visa fler...
-
- Hult, J (författare)
- Lund University,Lunds universitet,Oftalmologi, Lund,Sektion IV,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Forskargruppen för oftalmologisk avbildning,Forskargrupper vid Lunds universitet,Ophthalmology, Lund,Section IV,Department of Clinical Sciences, Lund,Faculty of Medicine,Ophthalmology Imaging Research Group,Lund University Research Groups,Skåne University Hospital
-
- Reistad, N (författare)
- Lund University,Lunds universitet,Atomfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Nationellt Resurscentrum för Fysik,Atomic Physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH,National Resource Centre for Physics Education,Faculty of Engineering, LTH
-
- Malmsjö, M (författare)
- Lund University,Lunds universitet,Oftalmologi, Lund,Sektion IV,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,NPWT teknologin,Forskargrupper vid Lunds universitet,Klinisk och experimentell lungtransplantation,Forskargruppen för oftalmologisk avbildning,Ophthalmology, Lund,Section IV,Department of Clinical Sciences, Lund,Faculty of Medicine,NPWT technology,Lund University Research Groups,Clinical and experimental lung transplantation,Ophthalmology Imaging Research Group,Skåne University Hospital
-
visa färre...
-
(creator_code:org_t)
- 2018-05-17
- 2018
- Engelska 5 s.
-
Ingår i: Skin Research and Technology. - : Wiley. - 0909-752X. ; 24:4, s. 667-671
- Relaterad länk:
-
http://dx.doi.org/10... (free)
-
visa fler...
-
https://onlinelibrar...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- OBJECTIVE: A novel extended-wavelength diffuse reflectance spectroscopy (EWDRS) technique is being developed for future clinical non-invasive tumor margin delineation. In this study, the ability of EWDRS to identify the margins of pigmented skin lesions in an in vivo pig model was evaluated.MATERIALS AND METHODS: Extended-wavelength diffuse reflectance spectroscopy recordings (350-1550 nm) were made on 13 pigmented skin lesions and non-pigmented skin, as a reference. The hand-held probe was swept toward the pigmented area until the signal changed, thus indicating that the margin had been identified. A needle was inserted as a marker, and tissue samples were sent for histological analysis. The distance between the EWDRS-defined border and the histological border was measured by 3 independent examiners.RESULTS: The median difference between the EWDRS-defined border and the histological border was 70 μm toward the pigmented tissue (range: -579 to 538 μm). A Pearson correlation coefficient of .95 was obtained for the examiners.CONCLUSIONS: Extended-wavelength diffuse reflectance spectroscopy can be used in vivo to delineate the border of pigmented skin lesions in a porcine model with high accuracy, indicating that it may be a useful tool for non-invasive tumor margin delineation in the future.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Oftalmologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Ophthalmology (hsv//eng)
- MEDICIN OCH HÄLSOVETENSKAP -- Medicinsk bioteknologi -- Biomedicinsk laboratorievetenskap/teknologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Medical Biotechnology -- Biomedical Laboratory Science/Technology (hsv//eng)
Nyckelord
- Journal Article
Publikations- och innehållstyp
- art (ämneskategori)
- ref (ämneskategori)
Hitta via bibliotek