Beam control system and output fine-tuning for safe and precise delivery of FLASH radiotherapy at a clinical linear accelerator

Konradsson, Elise (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Radiotherapy Physics,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Lund University Research Groups
Wahlqvist, Pontus (författare)
Skåne University Hospital
Thoft, Andreas (författare)
Skåne University Hospital
visa fler...
Blad, Börje (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups,Skåne University Hospital
Bäck, Sven (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Radiotherapy Physics,LUCC: Lunds universitets cancercentrum,Övriga starka forskningsmiljöer,Medical Radiation Physics, Malmö,Lund University Research Groups,LUCC: Lund University Cancer Centre,Other Strong Research Environments,Skåne University Hospital
Ceberg, Crister (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Radiotherapy Physics,Forskargrupper vid Lunds universitet,Rausinglaboratoriet i Lund - Tumörsektionen,LU profilområde: Ljus och material,Lunds universitets profilområden,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Lund University Research Groups,Rausing laboratory of Lund - Tumor section,LU Profile Area: Light and Materials,Lund University Profile areas
Petersson, Kristoffer (författare)
University of Oxford,Skåne University Hospital
visa färre...
 (creator_code:org_t)
2024
2024
Engelska.
Ingår i: Frontiers in Oncology. - 2234-943X. ; 14
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Introduction: We have previously adapted a clinical linear accelerator (Elekta Precise, Elekta AB) for ultra-high dose rate (UHDR) electron delivery. To enhance reliability in future clinical FLASH radiotherapy trials, the aim of this study was to introduce and evaluate an upgraded beam control system and beam tuning process for safe and precise UHDR delivery. Materials and Methods: The beam control system is designed to interrupt the beam based on 1) a preset number of monitor units (MUs) measured by a monitor detector, 2) a preset number of pulses measured by a pulse-counting diode, or 3) a preset delivery time. For UHDR delivery, an optocoupler facilitates external control of the accelerator’s thyratron trigger pulses. A beam tuning process was established to maximize the output. We assessed the stability of the delivery, and the independent interruption capabilities of the three systems (monitor detector, pulse counter, and timer). Additionally, we explored a novel approach to enhance dosimetric precision in the delivery by synchronizing the trigger pulse with the charging cycle of the pulse forming network (PFN). Results: Improved beam tuning of gun current and magnetron frequency resulted in average dose rates at the dose maximum at isocenter distance of >160 Gy/s or >200 Gy/s, with or without an external monitor chamber in the beam path, respectively. The delivery showed a good repeatability (standard deviation (SD) in total film dose of 2.2%) and reproducibility (SD in film dose of 2.6%). The estimated variation in DPP resulted in an SD of 1.7%. The output in the initial pulse depended on the PFN delay time. Over the course of 50 measurements employing PFN synchronization, the absolute percentage error between the delivered number of MUs calculated by the monitor detector and the preset MUs was 0.8 ± 0.6% (mean ± SD). Conclusion: We present an upgraded beam control system and beam tuning process for safe and stable UHDR electron delivery of hundreds of Gy/s at isocenter distance at a clinical linac. The system can interrupt the beam based on monitor units and utilize PFN synchronization for improved dosimetric precision in the dose delivery, representing an important advancement toward reliable clinical FLASH trials.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

beam control
clinical linear accelerator
electrons
FLASH-RT
ultra-high dose rate

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy