  • Martens, C (författare)

Underdosage of the upper-airway mucosa for small fields as used in intensity-modulated radiation therapy: A comparison between radiochromic film measurements, Monte Carlo simulations, and collapsed cone convolution calculations

  • Artikel/kapitelEngelska2002

Förlag, utgivningsår, omfång ...

  • 2002-06-21
  • Wiley,2002

Nummerbeteckningar

  • LIBRIS-ID:oai:lup.lub.lu.se:4c18ef91-780d-43b4-9136-05a2c9b53078
  • https://lup.lub.lu.se/record/333239URI
  • https://doi.org/10.1118/1.1487421DOI

Kompletterande språkuppgifter

  • Språk:engelska
  • Sammanfattning på:engelska

Ingår i deldatabas

Klassifikation

  • Ämneskategori:art swepub-publicationtype
  • Ämneskategori:ref swepub-contenttype

Anmärkningar

  • Head-and-neck tumors are often situated at an air-tissue interface what may result in an underdosage of part of the tumor in radiotherapy treatments using megavoltage photons. especially for small fields. In addition to effects of transient electronic disequilibrium, for these small fields, an increased lateral electron range in air will result in an important extra reduction of the central axis (lose beyond the cavity. Therefore dose calculation algorithms need to model electron transport accurately. We simulated the trachea by a 2 cm diameter cylindrical air cavity with the rin) situated 2 cm beneath the phantom surface. A 6 MV photon beam from an Elekta SLi plus linear accelerator, equipped with the standard multileaf collimator (MLC), was assessed. A 10 x 2 cm(2) and a 10 K 1 cm(2) field, both widthwise collimated by the MLC, were applied with their long side parallel to the cylinder axis. Central axis dose rebuild-up was studied. Radiochromic film measurements were performed in an in-house manufactured polystyrene phantom with the films oriented either along or perpendicular to the beam axis. Monte Carlo simulations were performed with BEAM and EGSnrc. Calculations were also performed using the pencil beam (PB) algorithm and the collapsed cone convolution (CCC) algorithm of Helax-TMS (MDS Nordion, Kanata. Canada) version 6.0.2 and using the CCC algorithm of Pinnacle (ADAC Laboratories, Milpitas. CA, USA) version 4.2. A very good agreement between the film measurements and the Monte Carlo simulations was found. The CCC algorithms were not able to predict the interface dose accurately when lateral electronic disequilibrium occurs, but were shown to be a considerable improvement compared to the PB algorithm. The CCC algorithms overestimate the dose in the rebuild-up region. The interface dose was overestimated by a maximum of 31% or 54%, depending on the implementation of the CCC algorithm. At a depth of I rum, the maximum dose overestimation was 14% or 24%.

Ämnesord och genrebeteckningar

Biuppslag (personer, institutioner, konferenser, titlar ...)

  • Reynaert, N (författare)
  • De Wagter, C (författare)
  • Nilsson, PerLund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine(Swepub:lu)rafy-per (författare)
  • Coghe, M (författare)
  • Palmans, H (författare)
  • Theirens, H (författare)
  • De Neve, W (författare)
  • Medicinsk strålningsfysik, LundSektion V (creator_code:org_t)

Sammanhörande titlar

  • Ingår i:Medical Physics: Wiley29:7, s. 1528-15350094-2405

Internetlänk

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy