Comparison of conventional and Si-photomultiplier-based PET systems for image quality and diagnostic performance

Oddstig, Jenny (författare)
Lund University,Lunds universitet,Nuklearmedicin, Malmö,Forskargrupper vid Lunds universitet,Nuclear medicine, Malmö,Lund University Research Groups,Skåne University Hospital
Leide Svegborn, Sigrid (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups,Skåne University Hospital
Almquist, Helen (författare)
Skåne University Hospital
visa fler...
Bitzén, Ulrika (författare)
Skåne University Hospital
Garpered, Sabine (författare)
Lund University,Lunds universitet,Klinisk fysiologi och nuklearmedicin, Malmö,Forskargrupper vid Lunds universitet,Clinical Physiology and Nuclear Medicine, Malmö,Lund University Research Groups,Skåne University Hospital
Hedeer, Fredrik (författare)
Skåne University Hospital
Hindorf, Cecilia (författare)
Lund University,Lunds universitet,Hjärt-MR-gruppen i Lund,Forskargrupper vid Lunds universitet,Lund Cardiac MR Group,Lund University Research Groups,Skåne University Hospital
Jögi, Jonas (författare)
Lund University,Lunds universitet,Hjärt-MR-gruppen i Lund,Forskargrupper vid Lunds universitet,Lund Cardiac MR Group,Lund University Research Groups,Skåne University Hospital
Jönsson, Lena (författare)
Lund University,Lunds universitet,Nuclear Medicine Physics,Forskargrupper vid Lunds universitet,Lund University Research Groups,Skåne University Hospital
Minarik, David (författare)
Lund University,Lunds universitet,Nuklearmedicin, Malmö,Forskargrupper vid Lunds universitet,Nuclear medicine, Malmö,Lund University Research Groups,Skåne University Hospital
Petersson, Richard (författare)
Skåne University Hospital
Welinder, Annika (författare)
Skåne University Hospital
Wollmer, Per (författare)
Lund University,Lunds universitet,Klinisk fysiologi och nuklearmedicin, Malmö,Forskargrupper vid Lunds universitet,Clinical Physiology and Nuclear Medicine, Malmö,Lund University Research Groups,Skåne University Hospital
Trägårdh, Elin (författare)
Lund University,Lunds universitet,Nuklearmedicin, Malmö,Forskargrupper vid Lunds universitet,WCMM- Wallenberg center för molekylär medicinsk forskning,Medicinska fakulteten,Nuclear medicine, Malmö,Lund University Research Groups,WCMM-Wallenberg Centre for Molecular Medicine,Faculty of Medicine,Skåne University Hospital
visa färre...
 (creator_code:org_t)
2019-10-22
2019
Engelska.
Ingår i: BMC Medical Imaging. - : Springer Science and Business Media LLC. - 1471-2342. ; 19:1
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • BACKGROUND: A new generation of positron emission tomography with computed tomography (PET-CT) was recently introduced using silicon (Si) photomultiplier (PM)-based technology. Our aim was to compare the image quality and diagnostic performance of a SiPM-based PET-CT (Discovery MI; GE Healthcare, Milwaukee, WI, USA) with a time-of-flight PET-CT scanner with a conventional PM detector (Gemini TF; Philips Healthcare, Cleveland, OH, USA), including reconstruction algorithms per vendor's recommendations. METHODS: Imaging of the National Electrical Manufacturers Association IEC body phantom and 16 patients was carried out using 1.5 min/bed for the Discovery MI PET-CT and 2 min/bed for the Gemini TF PET-CT. Images were analysed for recovery coefficients for the phantom, signal-to-noise ratio in the liver, standardized uptake values (SUV) in lesions, number of lesions and metabolic TNM classifications in patients. RESULTS: In phantom, the correct (> 90%) activity level was measured for spheres ≥17 mm for Discovery MI, whereas the Gemini TF reached a correct measured activity level for the 37-mm sphere. In patient studies, metabolic TNM classification was worse using images obtained from the Discovery MI compared those obtained from the Gemini TF in 4 of 15 patients. A trend toward more malignant, inflammatory and unclear lesions was found using images acquired with the Discovery MI compared with the Gemini TF, but this was not statistically significant. Lesion-to-blood-pool SUV ratios were significantly higher in images from the Discovery MI compared with the Gemini TF for lesions smaller than 1 cm (p < 0.001), but this was not the case for larger lesions (p = 0.053). The signal-to-noise ratio in the liver was similar between platforms (p = 0.52). Also, shorter acquisition times were possible using the Discovery MI, with preserved signal-to-noise ratio in the liver. CONCLUSIONS: Image quality was better with Discovery MI compared to conventional Gemini TF. Although no gold standard was available, the results indicate that the new PET-CT generation will provide potentially better diagnostic performance.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
NATURVETENSKAP  -- Fysik -- Annan fysik (hsv//swe)
NATURAL SCIENCES  -- Physical Sciences -- Other Physics Topics (hsv//eng)

Nyckelord

Digital PET
FDG
Image quality
Oncology
PET-CT

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy