Sealing of coronary perforations with a second-generation covered stent graft - Results from the PAST-PERF registry

Birkemeyer, Ralf (författare)
Herzklinik Ulm
Olivecrona, Göran K. (författare)
Lund University,Lunds universitet,Kardiologi,Sektion II,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Cardiology,Section II,Department of Clinical Sciences, Lund,Faculty of Medicine,Skåne University Hospital
Hellig, Farrel (författare)
Sunninghill Hospital
visa fler...
Wöhrle, Jochen (författare)
Clinical Center of Friedrichshafen,University Hospital of Ulm
Rottbauer, Wolfgang (författare)
University Hospital of Ulm
Witkowski, Adam (författare)
Cardinal Wyszynski National Institute of Cardiology
Kuliczkowski, Wiktor (författare)
Wrocław Medical University
Bernhardt, Peter (författare)
Herzklinik Ulm
Bettels, Nadine (författare)
Herzklinik Ulm
Schrage, Benedikt (författare)
University Medical Center Hamburg-Eppendorf
von zur Mühlen, Constantin (författare)
University Heart Center Freiburg Bad Krozingen
Cook, Stephane (författare)
Cantonal Hospital Fribourg
Miljak, Tomislav (författare)
Sindelfingen-Boeblingen Medical Center
Eggbrecht, Holger (författare)
Agaplesion Bethanien Hospital
Eeckhout, Eric (författare)
Lausanne University Hospital
Westermann, Dirk (författare)
University Medical Center Hamburg-Eppendorf
Monsegu, Jaques (författare)
Hospital Group Mutual De Grenoble
Dumonteil, Nicolas (författare)
Clinique Pasteur, Toulouse
visa färre...
 (creator_code:org_t)
Elsevier BV, 2021
2021
Engelska 7 s.
Ingår i: Cardiovascular Revascularization Medicine. - : Elsevier BV. - 1553-8389. ; 25, s. 20-26
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background: The PAST-PERF registry was initiated to collect data on the PK Papyrus covered stent, a second-generation device for the treatment of coronary artery perforations with enhanced mechanical properties, but with limited available data. Methods: Patients treated for coronary artery perforations with the PK Papyrus stent at 14 international centers were retrospectively identified. The primary effectiveness outcome was successful sealing of the perforation. The primary safety outcome was a composite of all-cause mortality, definite or probable stent thrombosis, myocardial infarction and target lesion revascularization. Results: Among the 94 included patients, 72.3% (68/94) had Ellis type III and cavity spilling perforations. Complete sealing was achieved in 93.6% (n = 88), and no sealing could be achieved in 3.2% (n = 3, including one patient with a geographical miss and one patient in whom the device could not be implanted). Pericardiocentesis was required in 25.0% (n = 23), emergency cardiac surgery was needed in 7.6% (n = 7), acute stent thrombosis was observed in 1.1% (n = 1), and in-hospital mortality occurred in 11.7% (n = 11). The median follow-up duration was 283 (IQR:40;670) days. At 6 and 12 months, the incidence of the primary safety endpoint was 26.6% [95%CI:18.6;37.1] and 32.0% [95%CI:22.8;43.4], mortality 15.0% [95%CI:9.0;24.6] and 19.0% [95%CI:11.3;30.0], and target lesion revascularization 5.5% [95%CI:2.0;14.6] and 7.7% [95%CI:3.1;18.2]. Two definite stent thrombosis occurred, one during the procedure and one on post-procedure day 233. Conclusions: The registry demonstrates favorably high rates of successful stent delivery and sealing of coronary perforations using a second-generation covered stent with low target lesion revascularization and stent thrombosis rates. Annotated table of content: The PAST-PERF registry demonstrates favorably high rates of successful stent delivery and sealing of coronary perforations using a second-generation covered stent with low target lesion revascularization and stent thrombosis rates. Specifically, complete sealing was achieved in 93.6% of patients (n = 88/94), and no sealing could be achieved in 3.2% (n = 3, including one patient with a geographical miss and one patient in whom the device could not be implanted). The 12-month mortality was 19.0% [95%CI:11.3;30.0], the rate of target lesion revascularization was 7.7% [95%CI:3.1;18.2], and two definite stent thromboses occurred (one during procedure and one on post-procedure day 233).

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Kardiologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Cardiac and Cardiovascular Systems (hsv//eng)

Nyckelord

Coronary artery disease
Coronary artery perforation
Covered stent
Percutaneous coronary intervention

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy