A telescopic stent-graft for aortoiliac implantation

Chuter, Timothy A M (author)
Malina, Martin (author)
Lund University,Lunds universitet,Institutionen för kliniska vetenskaper, Malmö,Medicinska fakulteten,Vaskulära sjukdomar - kliniska studier,Forskargrupper vid Lunds universitet,Department of Clinical Sciences, Malmö,Faculty of Medicine,Vascular Diseases - Clinical Research,Lund University Research Groups
Brunkwall, Jan (author)
show more...
Lindh, Mats (author)
Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Radiology Diagnostics, Malmö,Lund University Research Groups
Ivancev, Krassi (author)
Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Radiology Diagnostics, Malmö,Lund University Research Groups
Lindblad, Bengt (author)
Lund University,Lunds universitet,Vaskulära sjukdomar - kliniska studier,Forskargrupper vid Lunds universitet,Vascular Diseases - Clinical Research,Lund University Research Groups
Risberg, Bo (author)
show less...
 (creator_code:org_t)
1997
1997
English.
In: European Journal of Vascular and Endovascular Surgery. - 1532-2165. ; 13:1, s. 79-84
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • OBJECTIVES: To test a new delivery system for a telescopic stent-graft. MATERIALS: Variable overlap between the components of this stent-graft combination allows length adjustment. This device differs from other multi-component stent-grafts in that both components are contained within a single delivery system. METHODS: The stent-graft was implanted in the distal thoracic to suprarenal aorta of five pigs (35-50 kg), where the arterial diameter falls by almost 50%. The proximal and distal components of the stent-graft were targeted to bony landmarks in the vertebral column. RESULTS: Inspection of completion angiograms showed both proximal and distal stent-grafts to be within 1 mm of their target locations in all five experiments. Overall combined stent-graft length varied from 13.5 cm to 16.1 cm depending on the location of the bony landmark chosen as the distal target, and on the size of the pig. CONCLUSIONS: This system may be useful for the repair of abdominal aortic aneurysm whenever preoperative sizing is difficult due to aortic tortuosity, or precluded due to the urgency of the procedure.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Kirurgi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Surgery (hsv//eng)

Keyword

Endovascular
Aneurysm
Stent-graft
Telescopic

Publication and Content Type

art (subject category)
ref (subject category)

Find in a library

To the university's database

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view