Sökning: WFRF:(Mikko S) >
A high resolution A...
A high resolution AMS-injector for the Pelletron in Lund
-
- Hellborg, Ragnar (författare)
- Lund University,Lunds universitet,Kärnfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Nuclear physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH
-
Bazhal, S (författare)
-
- Faarinen, Mikko (författare)
- Lund University,Lunds universitet,Kärnfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Nuclear physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH
-
visa fler...
-
Hakansson, K (författare)
-
- Magnusson, Carl-Erik (författare)
- Lund University,Lunds universitet,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Physics,Departments at LTH,Faculty of Engineering, LTH
-
- Persson, Per (författare)
- Lund University,Lunds universitet,Kärnfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Nuclear physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH
-
- Skog, Göran (författare)
- Lund University,Lunds universitet,Kvartärgeologi,Geologiska institutionen,Naturvetenskapliga fakulteten,Quaternary Sciences,Department of Geology,Faculty of Science
-
- Stenström, Kristina (författare)
- Lund University,Lunds universitet,Kärnfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Nuclear physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH
-
visa färre...
-
(creator_code:org_t)
- 2002
- 2002
- Engelska.
-
Ingår i: Pramana. - 0973-7111. ; 59:6, s. 1061-1073
- Relaterad länk:
-
http://www.ias.ac.in...
-
visa fler...
-
https://lup.lub.lu.s...
-
visa färre...
Abstract
Ämnesord
Stäng
- A high resolution injector system has recently been installed at the Lund 3 MV tandem Pelletron accelerator. The new injector, designed mainly for Al-26 ions, will increase the experimental potential of the Lund AMS facility considerably. High quality energy- and mass-resolution is obtained by using a 90degrees spherical electrostatic analyzer followed by a 90degrees magnetic analyzer. The injector is equipped with a high intensity sputtering source with a spherical ionizer. A new analytical technique for acceptance calculations as well as PC-based computational methods have been used in the design of the ion optical system of the new injector. Compared to our old injector system which has a magnetic analyzer with a bending angle of only 15degrees, the new system has a more than ten times better resolution. The beam optics of the new system is also better designed to match the accelerator acceptance. In this way the ion transmission from the ion source to the detector, for different ions of interest in our AMS programme, has been increased.
Ämnesord
- NATURVETENSKAP -- Fysik -- Subatomär fysik (hsv//swe)
- NATURAL SCIENCES -- Physical Sciences -- Subatomic Physics (hsv//eng)
- NATURVETENSKAP -- Geovetenskap och miljövetenskap -- Geologi (hsv//swe)
- NATURAL SCIENCES -- Earth and Related Environmental Sciences -- Geology (hsv//eng)
Nyckelord
- AMS
- injector
- ion optical system
- electrostatic accelerator
- Pelletron
- accelerator mass spectrometry
Publikations- och innehållstyp