SwePub
Sök i LIBRIS databas

  Utökad sökning

WFRF:(Toril S)
 

Sökning: WFRF:(Toril S) > NT-proBNP on Cobas ...

NT-proBNP on Cobas h 232 in point-of-care testing: Performance in the primary health care versus in the hospital laboratory

Gils, Charlotte (författare)
Ramanathan, Ramshanker (författare)
Breindahl, Torben (författare)
visa fler...
Brokner, Mette (författare)
Christiansen, Anne L. (författare)
Eng, Oyvin (författare)
Hammer, Ida J. (författare)
Herrera, Catherine B. (författare)
Jansen, Aina (författare)
Langsjoen, Eva C. (författare)
Lokkebo, Elisabeth S. (författare)
Osestad, Toril (författare)
Schroder, Anne D. (författare)
Walther, Lisa (författare)
Lund University,Lunds universitet,Avdelningen för klinisk kemi och farmakologi,Institutionen för laboratoriemedicin,Medicinska fakulteten,Division of Clinical Chemistry and Pharmacology,Department of Laboratory Medicine,Faculty of Medicine
visa färre...
 (creator_code:org_t)
2015-08-11
2015
Engelska.
Ingår i: Scandinavian Journal of Clinical & Laboratory Investigation. - : Informa UK Limited. - 1502-7686 .- 0036-5513. ; 75:7, s. 602-609
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background. NT-proBNP may be useful for ruling out heart failure in primary health care. In this study we examined the analytical quality of NT-proBNP in primary health care on the Cobas h 232 point-of-care instrument compared with measurements performed in a hospital laboratory. Materials and methods. Blood samples requested for NT-proBNP were collected in primary health care (n = 95) and in a hospital laboratory (n = 107). NT-proBNP was measured on-site on Cobas h 232 instruments both in primary health care centres and at the hospital laboratory and all samples were also analyzed with a comparison method at the hospital. Precision, trueness, accuracy, and lot-variation were determined at different concentration levels and evaluated according to acceptance criteria. Furthermore user-friendliness was assessed by questionnaires. Results. For Cobas h 232 repeatability CV was 8.5-10.7% in the hospital setting and 5.3-10.0% in the primary health care and within the analytical quality specifications, but higher than with the comparison method (<4%). NT-proBNP results obtained in primary health care were significantly higher than by the hospital comparison method (bias ranged from 14.3-23.7%), whereas there was no significant bias when Cobas h 232 was used in the hospital setting (bias ranged from 4.9 to 7.0%). User-friendliness of Cobas h 232 was overall acceptable. Conclusion. Cobas h 232 point-of-care instrument for measurement of NT-proBNP performed satisfactorily with regard to precision, user-friendliness, and lot-variation. A decrease in NT-proBNP levels observed in samples transported to a central laboratory needs further attention and investigation.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Hälsovetenskap -- Hälso- och sjukvårdsorganisation, hälsopolitik och hälsoekonomi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Health Sciences -- Health Care Service and Management, Health Policy and Services and Health Economy (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Kardiologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Cardiac and Cardiovascular Systems (hsv//eng)

Nyckelord

Natriuretic peptide
biological markers
heart disease
point-of-care
systems
validation studies

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy