The Influence of Different Technique Factors on Image Quality for Chest Radiographs: Application of the Recent CEC Image Quality Criteria

Lanhede, B. (författare)
Tingberg, Anders (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Månsson, L.G. (författare)
visa fler...
Kheddache, S. (författare)
Widell, M. (författare)
Björneld, L. (författare)
Sund, P. (författare)
Almén, A. (författare)
Besjakov, Jack (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Radiology Diagnostics, Malmö,Lund University Research Groups
Mattsson, Sören (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Zankl, M. (författare)
Panzer, W. (författare)
Herrmann, C. (författare)
visa färre...
 (creator_code:org_t)
2000
2000
Engelska.
Ingår i: Radiation Protection Dosimetry. - 1742-3406. ; 90:1-2, s. 203-206
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The aim of the first this work part of the EU-project, Trial I, was to evaluate and possibly improve the CEC image criteria for radiographic chest images. Chest images of healthy volunteers were acquired using different technique factors. The image criteria were used as a tool to discriminate between the different images. The technique factors were chosen so that the image quality would differ slightly. Four different technique parameters, each with two possible settings, used in clinical practice today, were used: tube voltage - 102 and 141 kV; screen/film speed - 160 and 320; maximum optical density in the parenchyma - 1.3 and 1.8; method for scatter reduction - air gap 30/390 and moving grid40/12. The results showed that the image criteria were able to separate between different technique groups. Some conclusions can be drawn from the results Optical density 1.8 was better than 1.3 independent of the other parameters. . Among the six combinations ranked best , four used tube voltage 141 kV and four used air gap technique for scatter reduction. No difference was seen for screen/film speed. No correlation was seen between the ranking of the systems and patient dose.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy