Search: WFRF:(Dahlin Magnus) >
4,4′-Thiobis(2-tert...
4,4′-Thiobis(2-tert-butyl-5-methylphenol), an antioxidant in medical devices that may cause allergic contact dermatitis
-
- Antelmi, Annarita (author)
- Lund University,Lunds universitet,Yrkes- och miljödermatologi,Forskargrupper vid Lunds universitet,Occupational and Environmental Dermatology,Lund University Research Groups,Skåne University Hospital
-
- Lejding, Tina (author)
- Lund University,Lunds universitet,Yrkes- och miljödermatologi,Forskargrupper vid Lunds universitet,Occupational and Environmental Dermatology,Lund University Research Groups,Skåne University Hospital
-
- Bruze, Magnus (author)
- Lund University,Lunds universitet,Yrkes- och miljödermatologi,Forskargrupper vid Lunds universitet,Occupational and Environmental Dermatology,Lund University Research Groups,Skåne University Hospital
-
show more...
-
- Mowitz, Martin (author)
- Lund University,Lunds universitet,Yrkes- och miljödermatologi,Forskargrupper vid Lunds universitet,Occupational and Environmental Dermatology,Lund University Research Groups,Skåne University Hospital
-
- Dahlin, Jakob (author)
- Lund University,Lunds universitet,Yrkes- och miljödermatologi,Forskargrupper vid Lunds universitet,Occupational and Environmental Dermatology,Lund University Research Groups,Skåne University Hospital
-
show less...
-
(creator_code:org_t)
- 2023
- 2023
- English 4 s.
-
In: Contact Dermatitis. - 0105-1873. ; 89:2, s. 103-106
- Related links:
-
http://dx.doi.org/10...
-
show more...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
show less...
Abstract
Subject headings
Close
- Background: Antioxidants in medical devices, added to protect polymers or adhesives, may also cause contact dermatitis in some individuals. Objectives: To present data on sensitization to 4,4′-thiobis(2-tert-butyl-5-methylphenol), an antioxidant detected in some types of medical devices, for six patients that experienced eczematous reactions to different medical devices. Methods: Patch testing with 4,4′-thiobis(2-tert-butyl-5-methylphenol), 1% pet was performed. Gas chromatography-mass spectrometry (GC-MS) was used for identification of 4,4′-thiobis(2-tert-butyl-5-methylphenol) in different medical device products. Results: Six patients with contact allergy to 4,4′-thiobis(2-tert-butyl-5-methylphenol) also had relevant contact allergic reactions to medical devices containing the antioxidant. The presence of the antioxidant in products was detected using GC-MS analysis. Conclusions: The antioxidant 4,4′-thiobis(2-tert-butyl-5-methylphenol) may cause allergic contact dermatitis after exposure to different medical devices.
Subject headings
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Dermatologi och venereologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Dermatology and Venereal Diseases (hsv//eng)
Keyword
- antioxidant
- CAS 96-69-5
- contact dermatitis
- EMLA
- GC-MS
- Kendall AMD
- Lowinox
- medical device
- Santonox
- Steri-strip
- Tegaderm
- wound care
Publication and Content Type
- art (subject category)
- ref (subject category)
Find in a library