Evaluation of the image quality of ink-jet printed paper copies of digital chest radiographs as compared with film : A receiver operating characteristic study
-
- Lyttkens, Kerstin (författare)
- Lund University,Lunds universitet,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Skåne University Hospital
-
- Kirkhorn, Tomas (författare)
- Lund University,Lunds universitet,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medicinsk teknik, Lund,Institutionen för kliniska vetenskaper, Lund,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Biomedical Engineering, Lund,Department of Clinical Sciences, Lund
-
- Kehler, Mikael (författare)
- Skåne University Hospital,Lund University
-
visa fler...
-
Andersson, Berth (författare)
-
- Ebbesen, Anders (författare)
- Skåne University Hospital,Lund University
-
- Hochbergs, Peter (författare)
- Lund University,Lunds universitet,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Skåne University Hospital
-
- Jarlman, Olof (författare)
- Lund University,Lunds universitet,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
- Lindberg, Claes Göran (författare)
- Skåne University Hospital,Lund University
-
- Holmer, Nils Gunnar (författare)
- Lund University,Lunds universitet,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medicinsk teknik, Lund,Institutionen för kliniska vetenskaper, Lund,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Biomedical Engineering, Lund,Department of Clinical Sciences, Lund
-
visa färre...
-
(creator_code:org_t)
- 1994
- 1994
- Engelska.
-
Ingår i: Journal of Digital Imaging. - 0897-1889. ; 7:2, s. 61-68
- Relaterad länk:
-
http://dx.doi.org/10...
-
visa fler...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Paper copies of digital radiographs printed with the continuous ink-jet technique have proved to be of a high enough quality for demonstration purposes. We present a study on the image quality of ink-jet printed paper copies of digital chest radiographs, based on receiver operating characteristic (ROC) analysis. Eighty-three digital radiographs of a chest phatom with simulated tumors in the mediastinum and right lund, derived from a computed radiography (CR) system were presented in two series of hard copies as ink-jet printed paper copies and as laser recorded film. The images, with a matrix of 1,760×2,140 pixels, were printed with a spatial resolution of 10 pixels/mm in the CR film recorder as well as in the ink-jet printer. On film, every image was recorded in two versions, one optimized for the mediastinum and one for the lungs. On paper, only one image was printed; this constituted an effort to optimize both the mediastinum and the lungs. The ink-jet printed images, printed on a matt coated paper, were viewed as on-sight images with reflected light. The exdaminations were reviewed by six radiologists, and ROC curves were constructed. No significant difference was found between the performance of film and that of ink-jet paper prints. Because the cost for a paper copy is only a tenth of that of film, remarkable cost reductions can be achieved by using the ink jet technique instead. Our results show that further quality studies of ink-jet printed images are worthwhile.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Nyckelord
- hard copy
- ink-jet printing
- medical imaging
- phantom
- radiogrpahy digital
- receiver operating characteristics (ROC)
- thorax