SwePub
Sök i LIBRIS databas

  Extended search

L773:1879 0887 OR L773:0167 8140
 

Search: L773:1879 0887 OR L773:0167 8140 > (2005-2009) > Photon pencil kerne...

Photon pencil kernel parameterisation based on beam quality index

Nyholm, Tufve (author)
Umeå universitet,Radiofysik
Olofsson, Jörgen (author)
Umeå universitet,Radiofysik
Ahnesjö, Anders (author)
Umeå universitet,Uppsala universitet,Avdelningen för sjukhusfysik,Radiofysik
show more...
Karlsson, Mikael (author)
Umeå universitet,Radiofysik
show less...
 (creator_code:org_t)
Elsevier BV, 2006
2006
English.
In: Radiotherapy and Oncology. - : Elsevier BV. - 0167-8140 .- 1879-0887. ; 78:3, s. 347-351
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • Background and purposeNew treatment techniques in radiotherapy employ increasing dose calculation complexity in treatment planning. For an adequate check of the results coming from a modern treatment planning system, clinical tools with almost the same degree of generality and accuracy as the planning system itself are needed. To fulfil this need we propose a photon pencil kernel parameterization based on a minimum of input data that can be used for phantom scatter calculations. Through scatter integration the pencil kernel model can calculate common parameters, such as TPR or phantom scatter factors, used in various dosimetric QA (quality assurance) procedures.Material and methodsThe proposed model originates from an already published radially parameterized pencil kernel. A depth parameterization of the pencil kernel parameters has been introduced, based on a large database containing commissioned beam data for a commercial treatment planning system. The entire pencil kernel model demands only one photon beam quality index, TPR20,10, as input.ResultsBy comparing the dose calculation results to the extensive experimental data set in the database, it has been possible to make a thorough analysis of the resulting accuracy. The errors in calculated doses, normalized to the reference geometry, are in most cases smaller than 2%.ConclusionsThe investigation shows that a pencil kernel model based only on TPR20,10 can be used for dosimetric verification purposes in megavoltage photon beams at depths below the range of contaminating electrons.

Keyword

Algorithms
Computer Simulation
Models; Biological
Photons/*therapeutic use
Quality Assurance; Health Care/methods
Quality Control
Radiometry/*methods
Radiotherapy Dosage
Radiotherapy Planning; Computer-Assisted/*methods
Scattering; Radiation

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library