SwePub
Sök i LIBRIS databas

  Utökad sökning

WFRF:(Ullman Anders)
 

Sökning: WFRF:(Ullman Anders) > Comparison of clini...

  • Sandborg, MichaelLinköpings universitet,Medicinsk radiofysik,Hälsouniversitetet (författare)

Comparison of clinical and physical measures of image quality in chest and pelvis computed radiography at different tube voltages

  • Artikel/kapitelEngelska2006

Förlag, utgivningsår, omfång ...

  • 2006-10-18
  • Wiley,2006

Nummerbeteckningar

  • LIBRIS-ID:oai:lup.lub.lu.se:dacc12c4-f807-4b39-8161-20294d2dfb5c
  • https://lup.lub.lu.se/record/685772URI
  • https://doi.org/10.1118/1.2362871DOI
  • https://urn.kb.se/resolve?urn=urn:nbn:se:liu:diva-13220URI

Kompletterande språkuppgifter

  • Språk:engelska
  • Sammanfattning på:engelska

Ingår i deldatabas

Klassifikation

  • Ämneskategori:art swepub-publicationtype
  • Ämneskategori:ref swepub-contenttype

Anmärkningar

  • The aim of this work was to study the dependence of image quality in digital chest and pelvis radiography on tube voltage, and to explore correlations between clinical and physical measures of image quality. The effect on image quality of tube voltage in these two examinations was assessed using two methods. The first method relies on radiologists' observations of images of an anthropomorphic phantom, and the second method was based on computer modeling of the imaging system using an anthropomorphic voxel phantom. The tube voltage was varied within a broad range (50-150 kV), including those values typically used with screen-film radiography. The tube charge was altered so that the same effective dose was achieved for each projection. Two x-ray units were employed using a computed radiography (CR) image detector with standard tube filtration and antiscatter device. Clinical image quality was assessed by a group of radiologists using a visual grading analysis (VGA) technique based on the revised CEC image criteria. Physical image quality was derived from a Monte Carlo computer model in terms of the signal-to-noise ratio, SNR, of anatomical structures corresponding to the image criteria. Both the VGAS (visual grading analysis score) and SNR decrease with increasing tube voltage in both chest PA and pelvis AP examinations, indicating superior performance if lower tube voltages are employed. Hence, a positive correlation between clinical and physical measures of image quality was found. The pros and cons of using lower tube voltages with CR digital radiography than typically used in analog screen-film radiography are discussed, as well as the relevance of using VGAS and quantum-noise SNR as measures of image quality in pelvis and chest radiography.

Ämnesord och genrebeteckningar

Biuppslag (personer, institutioner, konferenser, titlar ...)

  • Tingberg, AndersLund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups,Department of Radiation Physics, Malmö University Hospital, Malmö, Sweden(Swepub:lu)rfa-ati (författare)
  • Ullman, GustafLinköpings universitet,Medicinsk radiofysik,Hälsouniversitetet(Swepub:liu)gusul07 (författare)
  • Dance, David R.Joint Department of Physics, The Royal Marsden NHS Trust and Institute of Cancer Research, London (författare)
  • Alm Carlsson, GudrunLinköpings universitet,Medicinsk radiofysik,Hälsouniversitetet(Swepub:liu)gudal03 (författare)
  • Linköpings universitetMedicinsk radiofysik (creator_code:org_t)

Sammanhörande titlar

  • Ingår i:Medical Physics: Wiley33:11, s. 4169-41750094-2405

Internetlänk

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy