The influence of wavelength and probe configuration on findings of a skin vasoconstriction test when using laser Doppler perfusion devices.
-
- Freccero, Carolin (författare)
- Lund University,Lunds universitet,Institutionen för kliniska vetenskaper, Malmö,Medicinska fakulteten,Kirurgi,Forskargrupper vid Lunds universitet,Department of Clinical Sciences, Malmö,Faculty of Medicine,Surgery,Lund University Research Groups
-
- Wollmer, Per (författare)
- Lund University,Lunds universitet,Klinisk fysiologi och nuklearmedicin, Malmö,Forskargrupper vid Lunds universitet,Clinical Physiology and Nuclear Medicine, Malmö,Lund University Research Groups
-
- Sundkvist, Göran (författare)
- Lund University,Lunds universitet,Institutionen för kliniska vetenskaper, Malmö,Medicinska fakulteten,Department of Clinical Sciences, Malmö,Faculty of Medicine
-
visa fler...
-
- Svensson, Henry (författare)
- Lund University,Lunds universitet,Kirurgi,Forskargrupper vid Lunds universitet,Surgery,Lund University Research Groups
-
visa färre...
-
(creator_code:org_t)
- Elsevier BV, 2006
- 2006
- Engelska.
-
Ingår i: Microvascular Research. - : Elsevier BV. - 1095-9319 .- 0026-2862. ; 71:Jan 3, s. 64-67
- Relaterad länk:
-
http://www.ncbi.nlm....
-
visa fler...
-
http://dx.doi.org/10...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- The aim of this study was to establish the degree to which a standardized test based on laser Doppler blood flow measurement is dependent on the particular equipment set-up being used. For this purpose, we examined finger skin blood flow with laser Doppler instruments in 20 healthy subjects. In laser Doppler perfusion monitoring (LDPM), we used a custom-made probe with two detecting fibers placed 0.25 and 1.2 min from the illuminating fiber, respectively, and two laser Doppler perfusion imagers (LDPI) with a wavelength of 632.8 nm and 780 rim, respectively. Warming of the hand was achieved with a Peltier element, and reflex vasoconstriction was induced by immersing the other hand for 3 min into a water bath kept at 15 degrees C. As a measure for the change in skin blood flow, a vasoconstriction index (VAC: cooling/before cooling) was calculated and used for the comparison of the different devices. VAC values gathered around 0.6 for all devices. However, LDPI with a wavelength of 632.9 nm showed a slightly higher VAC index, and the difference was significant. We conclude that using a standardized test is the most appropriate for monitoring changes in blood flow rather than recording and comparing discrete values in intermittent recordings. Although a difference was noted when comparing the devices, different fiber separations and wavelengths seem then to be of little consequence. (c) 2005 Published by Elsevier Inc.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Kirurgi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Surgery (hsv//eng)
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Endokrinologi och diabetes (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Endocrinology and Diabetes (hsv//eng)
Nyckelord
- vasoconstriction
- skin blood flow
- probe configuration
- waveleng-th
- laser Doppler perfusion monitoring
- laser Doppler perfusion imaging
- sympathetic