SwePub
Sök i LIBRIS databas

  Utökad sökning

WFRF:(Stoops Erik)
 

Sökning: WFRF:(Stoops Erik) > (2015-2019) > Optimized Standard ...

Optimized Standard Operating Procedures for the Analysis of Cerebrospinal Fluid Aβ42 and the Ratios of Aβ Isoforms Using Low Protein Binding Tubes

Vanderstichele, Hugo Marcel Johan (författare)
ADx NeuroSciences
Janelidze, Shorena (författare)
Lund University,Lunds universitet,Klinisk minnesforskning,Forskargrupper vid Lunds universitet,Clinical Memory Research,Lund University Research Groups
Demeyer, Leentje (författare)
ADx NeuroSciences
visa fler...
Coart, Els (författare)
International Drug Development Institute
Stoops, Erik (författare)
ADx NeuroSciences
Herbst, Victor (författare)
Euroimmun Medizinische Labordiagnostika AG
Mauroo, Kimberley (författare)
ADx NeuroSciences
Brix, Britta (författare)
Hansson, Oskar (författare)
Lund University,Lunds universitet,Klinisk minnesforskning,Forskargrupper vid Lunds universitet,Clinical Memory Research,Lund University Research Groups,Skåne University Hospital
visa färre...
 (creator_code:org_t)
2016
2016
Engelska 12 s.
Ingår i: Journal of Alzheimer's Disease. - 1387-2877. ; 53:3, s. 1121-1132
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background: Reduced cerebrospinal fluid (CSF) concentration of amyloid-β1-42 (Aβ1-42) reflects the presence of amyloidopathy in brains of subjects with Alzheimer's disease (AD). Objective: To qualify the use of Aβ1-42/Aβ1-40 for improvement of standard operating procedures (SOP) for measurement of CSF Aβ with a focus on CSF collection, storage, and analysis. Methods: Euroimmun ELISAs for CSF Aβ isoforms were used to set up a SOP with respect to recipient properties (low binding, polypropylene), volume of tubes, freeze/thaw cycles, addition of detergents (Triton X-100, Tween-20) in collection or storage tubes or during CSF analysis. Data were analyzed with linear repeated measures and mixed effects models. Results: Optimization of CSF analysis included a pre-wash of recipients (e.g., tubes, 96-well plates) before sample analysis. Using the Aβ1-42/Aβ1-40 ratio, in contrast to Aβ1-42, eliminated effects of tube type, additional freeze/thaw cycles, or effect of CSF volumes for polypropylene storage tubes. 'Low binding' tubes reduced the loss of Aβ when aliquoting CSF or in function of additional freeze/thaw cycles. Addition of detergent in CSF collection tubes resulted in an almost complete absence of variation in function of collection procedures, but affected the concentration of Aβ isoforms in the immunoassay. Conclusion: The ratio of Aβ1-42/Aβ1-40 is a more robust biomarker than Aβ1-42 toward (pre-) analytical interfering factors. Further, 'low binding' recipients and addition of detergent in collection tubes are able to remove effects of SOP-related confounding factors. Integration of the Aβ1-42/Aβ1-40 ratio and 'low-binding tubes' into guidance criteria may speed up worldwide standardization of CSF biomarker analysis.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Neurologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Neurology (hsv//eng)

Nyckelord

Amyloid
cerebrospinal fluid
ELISA
improvement
standard operating procedure

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy