Assessing TEG6S reliability between devices and across multiple time points: A prospective thromboelastography validation study

  • Artikel/kapitelEngelska2020

Förlag, utgivningsår, omfång ...

  • 2020-04-27
  • Springer Science and Business Media LLC,2020

Nummerbeteckningar

  • LIBRIS-ID:oai:prod.swepub.kib.ki.se:143639136
  • http://kipublications.ki.se/Default.aspx?queryparsed=id:143639136URI
  • https://doi.org/10.1038/s41598-020-63964-yDOI

Kompletterande språkuppgifter

  • Språk:engelska
  • Sammanfattning på:engelska

Ingår i deldatabas

Klassifikation

  • Ämneskategori:ref swepub-contenttype
  • Ämneskategori:art swepub-publicationtype

Anmärkningar

  • The TEG6S is a novel haemostasis analyser utilising resonance technology. It offers potentially greater coagulation information and ease of use, however has not been independently validated in a clinical setting. We aimed to determine if the TEG6S is reliable between devices and across time points. We performed a prospective observational study with ethical approval. For interdevice reliability, we performed simultaneous analysis on two TEG6S devices on 25 adult ICU patients. For time point reliability, we performed repeated sampling across five different time points on 15 adult participants. Blood was collected with informed consent, or as standard care, before four-channel citrated kaolin analysis. We observed almost perfect interdevice reliability across all TEG parameters. The Lin’s concordance correlation coefficients (95% CI, major axis regression slope, intercept) were R-time: 0.96 (0.92–0.99, 0.88, 0.57); K-time: 0.93 (0.87–0.98, 1.07, 0.00); Alpha Angle: 0.87 (0.78–0.96, 1.20, −14.10); Maximum Amplitude: 0.99 (0.98–0.99, 1.02, −1.38); Clot Lysis: 0.89 (0.82–0.97, 1.20, 0.07). Additionally, we observed moderate-to-high reliability across time points. Demonstrating almost perfect agreement across different devices and moderate-to-high reliability across multiple time points, suggests the TEG6S platform can be used with haemostatic accuracy and generalisability. This has potentially significant implications for clinical practice and multi-site research programs.

Biuppslag (personer, institutioner, konferenser, titlar ...)

  • Churilov, L (författare)
  • Cheong, B (författare)
  • Bellomo, R (författare)
  • McCall, PR (författare)
  • Martensson, JKarolinska Institutet (författare)
  • Glassford, N (författare)
  • Weinberg, L (författare)
  • Karolinska Institutet (creator_code:org_t)

Sammanhörande titlar

  • Ingår i:Scientific reports: Springer Science and Business Media LLC10:1, s. 7045-2045-2322

Internetlänk

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy