Do newborn infants exhale through the CPAP system? Secondary analysis of a randomised cross-over trial

Gunnarsdottir, K (författare)
Karolinska Institutet
Falk, M (författare)
Baldursdottir, S (författare)
Karolinska Institutet
visa fler...
Donaldsson, S (författare)
Karolinska Institutet
Jonsson, B (författare)
Karolinska Institutet
Drevhammar, T (författare)
Karolinska Institutet
visa färre...
 (creator_code:org_t)
2022-10-19
2023
Engelska.
Ingår i: Archives of disease in childhood. Fetal and neonatal edition. - : BMJ. - 1468-2052 .- 1359-2998. ; 108:3, s. 232-236
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • During nasal continuous positive airway pressure (nCPAP) treatment in neonates, leakage is inevitable and can lead to reduced distending pressure in the lungs of the infant. In current practice, neither leakage nor expiratory flow is measured, which makes it difficult to assess if exhalation is through the device or entirely through leakages.ObjectiveTo examine if infants treated with nCPAP exhale through the CPAP system.Design and settingSecondary data analyses from the ToNIL trial on leakages during nCPAP treatment. We retrospectively examined respiratory curves for the 50 infants included in the trial, using NI LabVIEW 2015. Each infant was measured with both prongs and nasal masks. A flow recording was classified as exhalation through the system if more than 50% of all expirations showed reverse flow, each for a minimum duration of 0.1 s.Patients50 infants were included, born with a mean gestational age (GA) of 34 weeks, median birth weight of 1948 g and mean age at measurement 6.5 days. Inclusion criteria were CPAP treatment and a postmenstrual age (PMA) of 28–42 weeks.ResultsIn our measurements, 32/50 infants exhaled through the CPAP system in at least one recording with either nasal mask or prongs. Leakages exceeding 0.3 L/min were seen in 97/100 recordings.ConclusionsDuring nCPAP treatment, infants can exhale through the CPAP system and leakage was common. Measuring expiratory flows and leakages in clinical settings could be valuable in optimising CPAP treatment of infants.Trial registration numberNCT03586856.

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy