Sökning: WFRF:(Håkansson Bo)
> (2020-2024) >
A novel method for ...
A novel method for objective in-situ measurement of audibility in bone conduction hearing devices - a pilot study using a skin drive BCD
-
- Persson, Ann-Charlotte, 1970 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för öron-, näs- och halssjukdomar,Institute of Clinical Sciences, Department of Otorhinolaryngology,Västra Götalandsregionen,Region Västra Götaland,University of Gothenburg
-
- Håkansson, Bo, 1953 (författare)
- Chalmers tekniska högskola,Chalmers University of Technology
-
- Mechanda, M. C. (författare)
- Chalmers tekniska högskola,Chalmers University of Technology
-
visa fler...
-
- Hodgetts, W. B. (författare)
- University of Alberta
-
- Fredén Jansson, Karl-Johan, 1988 (författare)
- Chalmers tekniska högskola,Chalmers University of Technology
-
- Eeg-Olofsson, Måns, 1967 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för öron-, näs- och halssjukdomar,Institute of Clinical Sciences, Department of Otorhinolaryngology,University of Gothenburg,Sahlgrenska universitetssjukhuset,Sahlgrenska University Hospital
-
- Reinfeldt, Sabine, 1978 (författare)
- Chalmers tekniska högskola,Chalmers University of Technology
-
visa färre...
-
(creator_code:org_t)
- 2022-03-03
- 2023
- Engelska.
-
Ingår i: International Journal of Audiology. - : Informa UK Limited. - 1499-2027 .- 1708-8186. ; 62:4, s. 357-361
- Relaterad länk:
-
https://www.tandfonl...
-
visa fler...
-
https://research.cha... (primary) (free)
-
https://gup.ub.gu.se...
-
https://doi.org/10.1...
-
https://research.cha...
-
visa färre...
Abstract
Ämnesord
Stäng
- Objective Objective measurement of audibility (verification) using bone conduction devices (BCDs) has long remained an elusive problem for BCDs. For air conduction hearing aids there are well-defined and often used objective methods, and the aim of this study is to develop an objective method for BCDs. Design In a novel setup for audibility measurements of bone-anchored hearing aid (BAHA) attached via a soft band, we used a skin microphone (SM) on the forehead measuring in-situ sound field thresholds, maximum power output (MPO) and international speech test signal (ISTS) responses. Study sample Five normal-hearing persons. Result Using the electrical output of SM it was possible to objectively measure the audibility of a skin drive BCD, presented as an eSPL-o-gram showing thresholds, MPO and ISTS response. Normalised eSPL-o-gram was verified against corresponding FL-o-grams (corresponding force levels from skull simulator and artificial mastoid (AM)). Conclusion The proposed method with the SM can be used for objective measurements of the audibility of any BCDs based on thresholds, MPO and speech response allowing for direct comparisons of hearing and BCD output on the same graph using an eSPL-o-gram. After normalisation to hearing thresholds, the audibility can be assessed without the need for complicated calibration procedures.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Oto-rhino-laryngologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Otorhinolaryngology (hsv//eng)
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Annan medicinteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Other Medical Engineering (hsv//eng)
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Medicinsk apparatteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Medical Equipment Engineering (hsv//eng)
Nyckelord
- Objective measurements
- verification
- bone conduction devices
- hearing
- rehabilitation