Technical design of a new bone conduction implant (BCI) system

Taghavi, Hamidreza, 1981 (author)
Chalmers tekniska högskola,Chalmers University of Technology
Håkansson, Bo, 1953 (author)
Chalmers tekniska högskola,Chalmers University of Technology
Reinfeldt, Sabine, 1978 (author)
Chalmers tekniska högskola,Chalmers University of Technology
show more...
Eeg-Olofsson, Måns, 1967 (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för öron-, näs- och halssjukdomar,Institute of Clinical Sciences, Department of Otorhinolaryngology,University of Gothenburg
Fredén Jansson, Karl-Johan, 1988 (author)
Chalmers tekniska högskola,Chalmers University of Technology
Håkansson, Emil, 1986 (author)
Chalmers tekniska högskola,Chalmers University of Technology
Nasri, Bayan (author)
Chalmers tekniska högskola,Chalmers University of Technology
show less...
 (creator_code:org_t)
2015-06-12
2015
English.
In: International Journal of Audiology. - : Informa UK Limited. - 1499-2027 .- 1708-8186. ; 54:10, s. 736-744
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • Objective: The objective of this study is to describe the technical design and verify the technical performance of a new bone conduction implant (BCI) system. Design: The BCI consists of an external audio processor and an implanted unit called the bridging bone conductor. These two units use an inductive link to communicate with each other through the intact skin in order to drive an implanted transducer. Study sample: In this study, the design of the full BCI system has been described and verified on a skull simulator and on real patients. Results: It was found that the maximum output force (peak 107 dB re 1 N) of the BCI is robust for skin thickness range of 2-8 mm and that the total harmonic distortion is below 8% in the speech frequency range for 70 dB input sound pressure level. The current consumption is 7.5 mA, which corresponds to 5-7 days use with a single battery. Conclusions: This study shows that the BCI is a robust design that gives a sufficiently high output and an excellent sound quality for the hearing rehabilitation of indicated patients.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Oto-rhino-laryngologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Otorhinolaryngology (hsv//eng)

Keyword

bone anchored hearing aid
Bone conduction implant
implanted transducer
inductive links
long-term implantation
long-term implantation

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view