Quality assurance guidelines for interstitial hyperthermia

Dobsicek Trefna, Hana, 1979 (författare)
Chalmers tekniska högskola,Chalmers University of Technology
Schmidt, M. (författare)
Universitatsklinik Erlangen und Medizinische Fakultat
Van Rhoon, G. C. (författare)
Erasmus MC Cancer Institute
visa fler...
Kok, H. P. (författare)
Universiteit Van Amsterdam,University of Amsterdam
Gordeyev, S. S. (författare)
Lamprecht, U. (författare)
Eberhard Karls Universität Tübingen,Eberhard Karls University of Tübingen
Marder, D. (författare)
Nadobny, J. (författare)
Charité Universitätsmedizin Berlin,Charité University Medicine Berlin
Ghadjar, Pirus (författare)
Charité Universitätsmedizin Berlin,Charité University Medicine Berlin
Abdel-Rahman, S. (författare)
Kukiełka, A. M. (författare)
Strnad, V. (författare)
Universitatsklinik Erlangen und Medizinische Fakultat
Hurwitz, M. D. (författare)
Vujaskovic, Z. (författare)
University of Maryland
Diederich, C. J. (författare)
University of California
Stauffer, P. R. (författare)
Crezee, J. (författare)
Universiteit Van Amsterdam,University of Amsterdam
visa färre...
 (creator_code:org_t)
2019-01-24
2019
Engelska.
Ingår i: International Journal of Hyperthermia. - : Informa UK Limited. - 0265-6736 .- 1464-5157. ; 36:1, s. 277-294
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Quality assurance (QA) guidelines are essential to provide uniform execution of clinical hyperthermia treatments and trials. This document outlines the clinical and technical consequences of the specific properties of interstitial heat delivery and specifies recommendations for hyperthermia administration with interstitial techniques. Interstitial hyperthermia aims at tumor temperatures in the 40–44 °C range as an adjunct to radiation or chemotherapy. The clinical part of this document imparts specific clinical experience of interstitial heat delivery to various tumor sites as well as recommended interstitial hyperthermia workflow and procedures. The second part describes technical requirements for quality assurance of current interstitial heating equipment including electromagnetic (radiative and capacitive) and ultrasound heating techniques. Detailed instructions are provided on characterization and documentation of the performance of interstitial hyperthermia applicators to achieve reproducible hyperthermia treatments of uniform high quality. Output power and consequent temperature rise are the key parameters for characterization of applicator performance in these QA guidelines. These characteristics determine the specific maximum tumor size and depth that can be heated adequately. The guidelines were developed by the ESHO Technical Committee with participation of senior STM members and members of the Atzelsberg Circle.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Maskinteknik -- Energiteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Mechanical Engineering -- Energy Engineering (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk apparatteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Equipment Engineering (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Samhällsbyggnadsteknik -- Husbyggnad (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Civil Engineering -- Building Technologies (hsv//eng)

Nyckelord

Clinical
quality assurance
hyperthermia
interstitial

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy