Comparisons of automated blood pressures in a primary health care setting with self-measurements at the office and at home using the Omron i-C10 device
-
- Al-Karkhi, Isam (författare)
- Linköpings universitet,Avdelningen för samhällsmedicin,Medicinska fakulteten
-
- Al-Rubaiy, Raad (författare)
- Linköpings universitet,Avdelningen för samhällsmedicin,Medicinska fakulteten
-
- Rosenqvist, Ulf (författare)
- Östergötlands Läns Landsting,Linköpings universitet,Avdelningen för kardiovaskulär medicin,Hälsouniversitetet,Medicinska specialistkliniken
-
visa fler...
-
- Falk, Magnus (författare)
- Östergötlands Läns Landsting,Linköpings universitet,Avdelningen för samhällsmedicin,Hälsouniversitetet,Primärvården i centrala länsdelen,Forsknings- och utvecklingsenheten för Närsjukvården i Östergötland
-
- Nyström, Fredrik H (författare)
- Östergötlands Läns Landsting,Linköpings universitet,Avdelningen för kardiovaskulär medicin,Hälsouniversitetet,Endokrinmedicinska kliniken
-
visa färre...
-
(creator_code:org_t)
- Lippincott Williams & Wilkins, 2015
- 2015
- Engelska.
-
Ingår i: Blood Pressure Monitoring. - : Lippincott Williams & Wilkins. - 1359-5237 .- 1473-5725. ; 20:2, s. 98-103
- Relaterad länk:
-
https://doi.org/10.1...
-
visa fler...
-
https://liu.diva-por... (primary) (Raw object)
-
https://urn.kb.se/re...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- OBJECTIVE: We aimed to compare blood pressure (BP) levels recorded using the semiautomatic oscillometric Omron i-C10 BP device in patients with or without hypertension in three different settings: (a) when used by a doctor or a nurse at the office (OBP); (b) when used for self-measurement by the patient at the office (SMOBP); and (c) when used for 7 consecutive days at home (HBP).MATERIALS AND METHODS: A total of 247 individuals were invited to participate, but 78 of these individuals declined and a further seven were excluded, leaving a final cohort of 162 participants.RESULTS: The mean OBP was higher than HBP (difference 8.1±14/3.1±8.8 mmHg, P<0.0001) and so was SMOBP compared with HBP (difference 7.0±13/4.2±7.3 mmHg, P<0.0001). Sixteen participants (9.9%) had at least 10 mmHg higher systolic SMOBP than OBP and 28 (17%) participants had at least 10 mmHg lower systolic SMOBP than OBP. Participants who were current smokers had a larger mean difference between systolic OBP and SMOBP than nonsmokers (OBP-SMOBP in smokers: 6.6±9.4 mmHg, OBP-SMOBP in nonsmokers: 0.5±9.2 mmHg, P=0.011 between groups).CONCLUSION: Self-measurement of BP in the office does not preclude an increase in BP when levels in the individual patients are compared with HBP using the same equipment. Thus, SMOBP with a semiautomatic device does not lead to a reduction in the white-coat effect in the same manner as fully automatic devices.This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License, where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially. http://creativecommons.org/licenses/by-nc-nd/3.0.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine (hsv//eng)
Nyckelord
- home blood pressure; hypertension; office blood pressure; oscillometric; smoking.
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas