5-ALA fluorescence and laser Doppler flowmetry for guidance in a stereotactic brain tumor biopsy

Haj-Hosseini, Neda, 1980- (author)
Linköpings universitet,Avdelningen för medicinsk teknik,Tekniska fakulteten,MINT
Richter, Johan, 1960- (author)
Linköpings universitet,Institutionen för medicinsk teknik,Tekniska fakulteten,Region Östergötland, Neurokirurgiska kliniken US
Milos, Peter, 1969- (author)
Linköpings universitet,Avdelningen för neuro- och inflammationsvetenskap,Medicinska fakulteten,Region Östergötland, Neurokirurgiska kliniken US,MINT
show more...
Hallbeck, Martin, 1970- (author)
Linköpings universitet,Avdelning för neurobiologi,Medicinska fakulteten,Region Östergötland, Klinisk patologi
Wårdell, Karin, 1959- (author)
Linköpings universitet,Avdelningen för medicinsk teknik,Tekniska fakulteten,MINT
show less...
 (creator_code:org_t)
Optical Society of America, 2018
2018
English.
In: Biomedical Optics Express. - : Optical Society of America. - 2156-7085. ; 9:5, s. 2284-2296
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • A fiber optic probe was developed for guidance during stereotactic brain biopsy procedures to target tumor tissue and reduce the risk of hemorrhage. The probe was connected to a setup for the measurement of 5-aminolevulinic acid (5-ALA) induced fluorescence and microvascular blood flow. Along three stereotactic trajectories, fluorescence (n = 109) and laser Doppler flowmetry (LDF) (n = 144) measurements were done in millimeter increments. The recorded signals were compared to histopathology and radiology images. The median ratio of protoporphyrin IX (PpIX) fluorescence and autofluorescence (AF) in the tumor was considerably higher than the marginal zone (17.3 vs 0.9). The blood flow showed two high spots (3%) in total. The proposed setup allows simultaneous and real-time detection of tumor tissue and microvascular blood flow for tracking the vessels.

Subject headings

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering (hsv//eng)

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view